product description
Fluidized bed granulation device
GB210-B
Used for granulation of powders and drying of wet powders
This device utilizes a flow layer for powder granulation and drying of wet powder. It is a flow layer drying and granulation device composed of the host unit GB210 and the attached GF200. It can correspond to 10000 energy specifications of AC200V, AC220V, and AC240V.
The experimental conditions for hot air temperature, air volume, adhesive transport liquid volume, etc. can be easily set through the setting knob on the front of the device
The drying chamber is made of super hard glass, which can observe the state of the flow layer and the state of spray. In addition, it is also very convenient to demonstrate the data of flowmeter, spray pressure gauge, inlet temperature and outlet temperature.
After purchasing GF3 spray accessories, it can not only spray but also flow bed granulation, with strong expansion, one machine can be used for many purposes, can cope with a variety of experiments, effectively save laboratory space, avoid multiple purchases, and save money.
Equipped with an electric lift for easy installation and disassembly of auxiliary devices.
200V240V multi standard power supply, touch screen operation, can switch between Chinese, English, and Japanese languages freely.
specifications
model |
|
GB210-B |
sexually
ability
|
Temperature regulator setting range |
40~220 ℃ (inlet temperature), 0-60 ℃ (outlet temperature) |
Temperature regulation accuracy |
Inlet temperature ± 1 ℃ |
Dry air volume adjustment range |
0~0.7m3/min |
Adjustment range of spray air pressure |
0~0.6MPa |
Flow range of liquid delivery pump |
0~26ml/min |
construct
accomplish
|
external output |
Inlet temperature and outlet temperature output (4-20mA) |
Automatic elevator |
The granulation component is automatically lifted and lowered by an elevator |
Temperature regulator |
PIDDigital temperature regulator |
Touch Screen |
Blower, heater, liquid delivery pump, switch for pulse jet, alarm display |
Control the toggle switch |
Switching between inlet temperature and outlet temperature control (outlet temperature control is conditional) |
temperature sensor |
KType thermocouple |
heater |
2.0KW(at200V)~2.88KW(at240V) |
Liquid delivery pump |
Pipe type liquid delivery pump |
Air pump for spray |
Use air compressor for spray (sold separately) |
Service socket |
Blender used: AC220V, 2A |
Inhalation blower |
Tube blower, brushless DC motor |
filter |
Inhalation filter, exhaust filter |
Cooling structure of spray nozzle |
Connection: connector × 2, outer diameter Φ 10.5mm |
Air connection for spray |
Outer diameter of the joint, Φ7mm |
Exhaust connection diameter |
Φ50mm |
Safety function |
Inlet and outlet temperature overheating, reverse function of liquid delivery pump, overcurrent leakage protection switch |
instrument for drawing circles
check
|
External dimensions |
W760×D420×H1350mm |
weight |
110kg |
Power supply (50/60Hz) |
AC200V 16A(AC220V 17A, AC240V 18A, wiring terminals need to be replaced) |
Accessories |
Liquid delivery hose (with 2 silicone stoppers and 3 TYLON stoppers), exhaust hose (with 1 stopper)
1 hose tie, outlet temperature sensor, fuse (250V 2A), anti-static wire, inlet
Air hose 5m (with 2 hose ties), GF300 spray assembly
|
shopping goods |
Safety protection cover (COV30C), temperature recorder, organic solvent recovery device (GAS410) |
control panel

Digital display of inlet temperature and outlet temperature
Temperature and dry air volume. And, through
Using touch screen for operation setting and operation
Switching status display and alarm display, can be arbitrary
Switch between Chinese, English, and Japanese, each can be set
Operating conditions.