Product Application: Widely used in the petrochemical industry to detect sulfur and chlorine content in liquid, solid, or gas samples.
Applicable standards: comply with SH/T0253 SH/T0254 SH/T0222 ASTMD3120 ASTMD3246, etc.
■ Functional features: WTL-06 microcoulometric titrator (microcomputer sulfur chlorine analyzer) is an application of microcoulometric analysis technology, using oxidation method
The sample is oxidized into titratable ions through a cracking furnace and titrated in a titration cell based on the amount of electricity consumed during the electrolytic titration process,
According to Faraday's law, calculate the content of sulfur or chlorine in the sample. The instrument has stable and reliable performance, easy operation, and analytical accuracy
High, good repeatability and other characteristics.
■ Technical parameters:
Measurement range: S: 0.2MG/L~5000 MG/L
N:0.2MG/L~5000 MG/L
CL:0.5MG/L~5000 MG/L
Current generated: ± 2MA
Electrolytic voltage: ± 30V
Detection limit: S: 0.2MG/L
N:0.2MG/L
CL:0.5MG/L
WTL-06 fully automatic microcoulometric sulfur and chlorine analyzer
Repetitive error: X<0.5MG/L, <50%
0.5MG/L≤X<1.0MG/L,±0.1MG/L
1.0MG/L≤X≤10MG/L,< 10%
X>10MG/L,<5%
Temperature control range: Room temperature~1000 ℃
Instrument usage environmental conditions: ambient temperature of 15-30 ℃, humidity ≤ 85%
Power supply voltage: AC220V, 50HZ
Dimensions: 1050 × 400 × 550MM (length × width × height)