Analytical Technology (Shenzhen) Co., Ltd
Home>Products>Total phosphorus detection reagent
Product Groups
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
  • Address
    4th Floor, Building C, Weijian Technology Industrial Park, Beihuan Road, Shiyan Street, Bao'an District, Shenzhen
Contact Now
Total phosphorus detection reagent
Total phosphorus detection reagent
Product details

Yuanyi Testing Brand Total Phosphorus Reagent(one)The total phosphorus reagent (II) and total phosphorus reagent (III) are composed of this reagent, which has the characteristics of simple configuration, stable and accurate detection, and easy use, avoiding configuration errors caused by insufficient experimental equipment for frontline testing personnel.

Technical parameters:


Reagents Name

Reagent specifications

Reagent composition

Storage method

Total phosphorus detection reagent

100order/cover

Reagent (1), Reagent (2), Reagent (3)

Refrigerated preservation/Store in a dark place

200order/cover

usage method:

1Prepare several clean and dry colorimetric tubes on the colorimetric tube rack.




2Add water sample to the colorimetric tube using a pipette or pipette.

2.1 Accurately measure5mlAdd distilled water to the colorimetric tube (zero point).

2.2 Accurately move them separately5mlTake each water sample and add it to its colorimetric tube in sequence (water sample).



3Add total phosphorus reagent (1) to each colorimetric tube in sequence, tighten the digestion tube cover and shake well. When the digestion instrument reaches the temperature, the buzzer will beep briefly to prompt. Insert the digestion tube into the digestion hole, cover it with a protective cover, and press the "confirm" button. The digestion instrument will start timing the digestion.


4. After the digestion is completed, the digestion instrument buzzer will sound a long beep, and pressing any key can cancel the long beep prompt. Place the digestion tube on the digestion tube rack and cool it to room temperature (either natural cooling or water cooling is acceptable).

5After cooling to room temperature, add sequentially0.2ml4Drip the total phosphorus reagent (II), cover and shake well, then let it stand30s(including blank space);Join in sequence0.3ml6Drip the total phosphorus reagent (III), cover and shake well, then let it stand for color development10 minutes, directly detect on the instrument.





Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!