1、 General Technical Parameters
Specification parameters
|
speak bright
|
power supply
|
9-45VDC.
|
output signal
|
Two line 4-20mA DC with digital communication (HART protocol) output.
|
Built in monitor (optional)
|
OLED displays numerical values, digits, units, and bar charts.
|
self-diagnosis
|
Amplifier failure, capsule failure, setting failure, environmental temperature alarm, pressure/static pressure alarm, etc.
|
operation temperature
|
-40-85 º C.
|
2、 Sensor type
1. SWP-S61DP series differential pressure transmitter
Sensor model
|
measuring range
|
Measurement range
|
Maximum working pressure
|
Accuracy/Stability
|
3
|
-7.5~7.5kPa
|
1.3~7.5kPa
|
7MPa
|
± 0.1% of the measuring range/
The maximum deviation per year is
± 0.1% of the measuring range
|
4
|
-37.4~37.4kPa
|
6.2~37.4 kPa
|
14MPa
|
5
|
-100~186.8kPa
|
31.1~186.8kPa
|
6
|
-100~690kPa
|
117~690kPa
|
7
|
-100~2068kPa
|
345~2068kPa
|
8
|
-100~6890kPa
|
1170~6890kPa
|
2. SWP-S61DR series micro differential pressure transmitter
Sensor model
|
measuring range
|
Measurement range
|
Maximum working pressure
|
Accuracy/Stability
|
2
|
-1.5~1.5kPa
|
0.125~1.5kPa
|
4MPa
|
± 0.1% of the measuring range/
The maximum deviation per year is
± 0.1% of the measuring range
|
3、SWP-ST61HPSeries high static pressure differential pressure transmitter
Sensor model
|
measuring range
|
Measurement range
|
Maximum working pressure
|
Accuracy/Stability
|
3
|
-7.5~7.5kPa
|
1.3~7.5kPa
|
25 MPa
32 MPa
40 Mpa
|
± 0.1% of the measuring range/
The maximum deviation per year is
± 0.1% of the measuring range
|
4
|
-37.4~37.4kPa
|
6.2~37.4 kPa
|
5
|
-100~186.8kPa
|
31.1~186.8kPa
|
6
|
-100~690kPa
|
117~690kPa
|
7
|
-100~2068kPa
|
345~2068kPa
|
4. SWP-S611AP/GP series absolute pressure/pressure transmitter
sensor
|
AP
|
GP
|
measuring range
|
Measurement range
|
Maximum working pressure
|
Accuracy/Stability
|
3
|
×
|
○
|
-7.5~7.5kPa
|
1.3~7.5kPa
|
Range upper limit pressure
|
± 0.1% of the measuring range/
The maximum deviation per year is
± 0.1% of the measuring range
|
4
|
○
|
○
|
-37.4~37.4kPa
|
6.2~37.4 kPa
|
5
|
○
|
○
|
-100~186.8kPa
|
31.1~186.8kPa
|
6
|
○
|
○
|
-100~690kPa
|
117~690kPa
|
7
|
○
|
○
|
-100~2068kPa
|
345~2068kPa
|
8
|
○
|
○
|
-100~6890kPa
|
1170~6890kPa
|
9
|
×
|
○
|
-100~20680kPa
|
3480~20680kPa
|
0
|
×
|
○
|
-100~41370kPa
|
6890~41370kPa
|
Note: Sensor model "○" indicates support; 'X' indicates that it is not supported
|