Product Introduction
Quantitative parallel concentrator
Quantitative parallel concentrator is a fast and convenient sample concentrator. It not only saves valuable fume hood resources in the laboratory, but also has convenient operation, unmanned operation, automatic notification of concentration completion, and short evaporation time, which can completely replace traditional concentration devices such as rotary evaporators and ordinary nitrogen blowers.
Performance advantages:
Eight/ten sample channels can work simultaneously and can be combined arbitrarily, starting and stopping at any time, adding or removing at any time, flexible and convenient
Can be used with 200mL and 50mL concentration cups, and both types of concentration cups can be used simultaneously, with a steaming and drying mode
Spiral nitrogen blowing, the position and angle of the nitrogen blowing tube can be easily adjusted to achieve maximum concentration speed
Automatically stop nitrogen blowing after opening the lid of the concentration chamber, and automatically return to the previous nitrogen blowing state after closing the lid of the concentration chamber. Humanized design, no additional operation required
The front of the concentrator has a visual window, and the concentrator room is equipped with lighting, making the concentration process visible
The instrument is sealed and environmentally friendly, with a built-in super strong exhaust system that does not require the use of fume hood space
■ Can automatically detect concentration endpoint, 1mL/0.5mL optional, stable and reliable
■ Water bath heating, room temperature -99.9 ℃
■ Water level exceeds the limit alarm and automatically stops heating
Gas pressure exceeds the limit alarm and automatically cuts off the airflow for convenience and safety
Real time display of current temperature, pressure, and remaining concentration time
The human-computer interaction interface adopts a touch screen, which is user-friendly and easy to control
Four working modes: Sensor, Time, Sensor&Time, Manual, can meet the sample concentration needs of different types of users
The nitrogen blowing tube is designed without welding process to avoid serious corrosion near the welding point due to long-term use
Function Introduction:
The human-computer interaction interface adopts a touch screen User friendly interface, easy to operate
Four working modes: Sensor, Time, Sensor&Time, Manual, can meet the sample concentration needs of different types of users
Spiral nitrogen blowing, nitrogen flow rate can be continuously adjusted, and the position and angle of the nitrogen blowing tube can be conveniently adjusted
Adjust to achieve maximum concentration speed (patent)
Equipped with a visual window and lighting, it can observe the concentrated state at any time without opening the cover Avoid frequent lid opening that affects concentration efficiency
Automatically stop nitrogen blowing after opening the lid of the concentration chamber, and automatically return to the previous nitrogen blowing state after closing the lid of the concentration chamber. Humanized design requires no additional operation
Nitrogen blowing tube without welding process design Avoid prolonged use as severe corrosion may occur near the solder joints
Eight/ten sample channels can work simultaneously and can be combined arbitrarily
■ Start stop at any time, add or remove at any time Flexible and convenient
Can be used with 200mL and 50mL concentration cups
Both types of concentration cups can be used simultaneously and have a steaming and drying mode
Multiple concentration cups are available, and the cup holder can be shared with fully automatic liquid processors such as GPC and SPE The sample does not need to be transferred and can be directly concentrated
Function Introduction:
The nitrogen flow rate of 8/10 groups can be adjusted separately, and the nitrogen flow rate control adopts an automatic proportional adjustment distribution system. The flow distribution is uniform, the flow rate is stable, and the flow rate can be freely set
Equipment parameters
PRODUCT MODEL |
CMX-10 |
CMX-10L |
temperature range |
Room temperature~99.9 ℃ |
Room temperature~99.9 ℃ |
precision |
0.1 ℃ |
0.1 ℃ |
Display mode |
LCD Touch Screen |
LCD Touch Screen |
Test tube capacity |
50/200ml |
50/200ml |
gas flow |
Adopting an automatic proportional adjustment distribution system, the flow distribution is uniform, the flow rate is stable, and the flow rate can be freely set |
Adopting an automatic proportional adjustment distribution system, the flow distribution is uniform, the flow rate is stable, and the flow rate can be freely set |
Water level alarm |
Water level exceeding alarm, automatic water injection, and automatic stop heating when there is no water |
Water level exceeding alarm, automatic water injection, and automatic stop heating when there is no water |
Heating method |
Constant temperature water bath |
Constant temperature water bath |
Sample size |
8 |
10/12 |
The nitrogen blowing needle can stretch up and down |
electric |
electric |
Fixed time/fixed volume stop |
Automatically turn off nitrogen gas [single group single control) |
Automatically turn off nitrogen gas [single group single control) |
|
Common solvent concentration rate (set temperature 60 ° C)
solvent |
Volume (mL) |
Approximate time to concentrate to 1mL (min) |
dichloromethane |
180 |
25 |
ethyl acetate |
180 |
20 |
cyclohexane |
180 |
18 |
methanol |
180 |
65 |
acetonitrile |
180 |
40 |
acetone |
180 |
21 |
hexane |
180 |
15 |
petroleum ether |
180 |
13 |