sensor |
sensing technology |
Michell ceramic humidity sensor |
Sensor version |
Easidew PRO I.S. |
range |
-100 ~ +20°Cdp |
Verify range |
-100 ~ +20°Cdp |
check |
Tracing back to the National Humidity Standards of the United Kingdom (NPL) and the United States (NIST) |
accuracy |
the dew point: ± 1 ° C at -60~+20 ° Cdp Moisture content: ± 10% of reading the dew point: ± 2 ° C at -60.1~-100 ° Cdp |
resolution |
0.1 ° C at+20~-100 ° Cdp |
Working pressure |
Up to 40 MPa |
operation temperature |
-40 ~ +60°C |
Sample flow rate |
1 ~ 5 l/min |
I. S. Certificate |
CSA certification IS Class I, Division 1, Groups A, B, C & D, T4 IECEx: Ex ia IIC T4 FM: IS/I/1/ABCD/T4 Ta=+60°C - Ex90385; Entity; IP66 |
Optional pressure sensor |
0-69 or 0-138 barg (other ranges are also available) Accuracy: ±0.25% FS |
Controller |
display |
Two rows of 6-digit LEDs display moisture content/dew point (user selected) and pressure |
Analog Output |
Two channels of 4-20 mA (maximum load 500 Ω) - user set parameters, units, and ranges |
Digital Output |
RS485 Modbus RTU |
display mode |
Moisture content (ppmV) Moisture content in natural gas (ppmV, lb/MMSCF, mg/m3) Dew point (° C or ° F) Temperature (° C or ° F) Pressure (PSI, barg) |
Pressure Compensation |
Fixed value (user set) or dynamic signal from pressure sensor |
display resolution |
0.1°Cdp, 0.1°Fdp, 0.1-0.001 ppmV ideal gas (optional), 0.01 ppmV natural gas, 0.01 mg/m3, 0.01 lb/MMSCF, 1 psig, 0.1 barg |
call the police |
Four alarm relays Trigger type and set point selectable Two way mode C contact level 10A, 240 V AC or 8 A, 24 V DC non inductive load Two way mode A contact level 5A, 240 V AC or 4 A, 24 V DC non inductive load |
I. S. Safety Barrier |
Isolation type, with built-in control module |
power supply |
85-265 V AC 50/60Hz or 10-72 V DC, maximum power consumption 10 W or 4 A, 24 V DC |
Internal cable |
Universal type, twisted pair shielded wire, single pair (or two pairs with pressure sensor) |
casing |
19 "rack control panel Size 132 x 483 x 256mm (h x w x d) (with a minimum clearance of 100mm at the rear end) |
work environment |
Indoor, safe area, 0 ~ +50°C, < 90% RH |
A comprehensive sampling system |
casing |
304 stainless steel (EN1.4301) cabinet; Can be customized into 316 stainless steel (EN1.4401) according to requirements; Made entirely of stainless steel material; Built in installation panel electroplating treatment; Indoor installation can choose open panel type Size 800 x 600 x 300mm (h x w x d) |
install |
Stainless steel wall mounted installation bracket |
protection grade |
IP66 |
temperature control |
Heater/temperature regulator option, can be a fixed control temperature point of+20 ° C or an adjustable temperature range of 0~50 ° C |
Heater power supply |
110/120 or 220/240/255 V AC, 50/60 Hz The maximum power consumption of the power supply is 100 W |
work environment |
Shaded areas, on land or at sea- When the ambient temperature is above+50 ° C (-40~+60 ° C max. transient), it is recommended to use the refrigeration option |