Oxygen analyzer suitable for use in hazardous environments
When an oxygen process application requires high precision and stability, as well as minimal maintenance costs, paramagnetic oxygen technology is usua
Product details

When an oxygen process application requires high precision and stability, as well as minimal maintenance costs, paramagnetic oxygen technology is usually prioritized. If the application also requires a robust, reliable, and corrosion-resistant analyzer, the latest XTP600 series thermal magnetic principle from Michell is the best choice.

highlight

  • Excellent accuracy of less than 1% of full scale ensures that the range can be reset without the need for further verification.
  • Excellent zero or range stability: 0.25% of the range per month
  • Innovative sensor design means zero drift and no background gas influence
  • Enhanced protection against pollution
  • Two 4-20 mA input signals are provided to the process variable for dynamic compensation
  • The two isolated outputs and local display inside the explosion-proof enclosure are suitable for both local and remote installation
  • Simplify maintenance to minimize user costs
XTP600 Series

technical indicators

performance
Thermomagnetic oxygen sensor
gas Process gas or non condensing gas
Measurement range Any oxygen content from 0~1% to 0~25% (special ranges, such as 0~100%, please consult first)
precision Better than ± 1% of the range
Response time (T90) 70 seconds. If you need faster, please consult.
repetitiveness ± 0.2% of the measuring range
linear Beyond the range of 0-5% is ± 0.5% of the measurement range
Within the range of 0-5%, it is better than ± 1% of the measuring range
Zero stability ± 0.25% of the measurement range per month
span stability ± 0.25% of the measurement range per month
Sample gas flow rate Sensing section 0.4 l/min
Maximum sample gas pressure Normal pressure to a maximum of 300 kPa
Maximum sample gas temperature Up to 45 ° C (please consult for higher temperatures first)
Electronic input/output
output signal Two 4~20 mA linear signals with safety isolation (one for oxygen and the other for background gas)
input Two 4~20 mA inputs for external background gas and pressure compensation
output range 0 ~ 1%; 2%; 5%; 10%; 25% (for higher ranges such as 0~100%, pressure compensation is required)
display resolution 0.1% within the range of 0-25%
power supply 24 V DC , Maximum current of 1.5 A
working conditions
ambient temperature 0°C ~ +45°C
Mechanical indicators
warm-up time 30 Minutes
size 213mm x 180mm x 140mm
weight 6.3kg
Sealing material AISI 316 and viton, other materials on request
Gas interface 1/4” NPT (f)
Protection level IP65。 Additional flame arresters are provided for hazardous applications
other
Dangerous Place Certificate II 2 GD Exd IIC T4 Gb II 2 D Extb IIIC T135°C Db IP66 T Amb -50°C/+60°C