Instrument function:
Conventional nucleic acid concentration testing:
1. Includes various nucleic acid samples such as dsDNA, ssDNA, RNA, etc., without the need for sample dilution
2. Automatically calculate and display 260nm and 280nm absorbance, ratio, and sample concentrationProtein detection:
1. Detect the concentration of purified protein and automatically adjust the optical path without diluting the sample
2. Detection range: 0.05-100mg/ul
Full spectrum scanning:
Full wavelength scanning from 1200 to 850nm, displaying absorption curve
2. Cell and Microbial Culture Test Substances
3. Detect the absorbance of fluorescent dye labeled proteins
Instrument features:
1. Only a small volume of sample is needed to reduce the required sample consumption, and no dilution is required. Samples ranging from 0.5ul to 2ul can be measured.
2. Detection wavelength range: 200-850nm
3. Extremely fast detection speed, no need for appointment, can be detected at any time. The measurement of each sample is completed in a very short time for testing
4. Longer optical lifespan, intelligent recognition of user usage, automatically turning off the light source within 5 minutes of no operation to extend its lifespan.
5. When the user presses the detection button, it will automatically turn on without user intervention, and the device will automatically recognize it
6. The graphical software interface is more intuitive, and the results can be directly exported in Excel spreadsheets for easy data viewing and analysis
Technical indicators and basic parameters
model |
N-100 |
Sample volume requirements |
0.5 - 2uL |
optical path |
0.2mm(High concentration measurement), 1.0mm(Ordinary concentration measurement) |
light source |
Pulse flashing xenon lamp |
detector |
3864Unit linear CCDarray |
wavelength range |
200-850nm |
wavelength accuracy |
±1nm |
Spectral bandwidth(Wavelength resolution) |
3nm(FWHMAt Hg 546nm) |
Photometric range |
0.02 - 75(Equivalent to 10mm)) |
photometric accuracy |
±0.003Abs, ± 1% T (0.76A at 350nm) |
Scope of nucleic acid testing |
4ng/uL - 5000ng/uL(dsDNA) |
Test time |
5second |
External dimensions |
W200 x D260 x H150mm |
weight |
2.5kg |
Sample holder material |
Quartz fiber and high hardness aluminum |
The power adapter |
DC 24V |
power consumption |
12-18W |
Data and Control Interface |
USBopening |
software platform |
Win XP(32Win 7 (32-bit) |