Product Code: |
111219121516 |
Product Name: |
Millipore ELIX Laboratory Pure Water System | |
Specifications: |
Elix Advantage | |
Product Remarks: |
||
Product Category: |
Ultrapure Water Polishing System | |
Elix AdvantageLaboratory pure water system technical specifications: ElixWater (in)ElixModule export) Parameter values resistivity>5MΩ.cm @ 25 °C TOC<30ppb If usedE-PODWater intake can achieve the following water quality indicators: parameter Numerical units bacteria<0.1 cfu/ml * particle>0.22 μm < 1 Particulates /ml* pyrogen(endotoxin) <0.001EU/ml ** RNases <0.01ng/ml ** DNases <4 pg/μl ** * Install withMillipore ExpressThe filter membraneMillipakFilter orBiopakIn the case of using ultrafiltration columns as terminal refiners. *During installationBiopakIn the case of using ultrafiltration columns as terminal refiners. integrate intoMerck MilliporeThe latest generation of professional technology in MilliporeElix AdvantageThe laboratory pure water system provides you with the best choice for pure water systems, while also elevating the quality of existing pure water in the market to new heights. In laboratory applications, the stability and reliability of water quality are crucial. Regulatory agencies have established minimum water quality requirements through specific and strict standards.Elix AdvantageThe water production of the system meets or exceeds the standardISO ® 3696 (2Grade water)、ASTM ® D1193 (IIGrade resistivity index andTOCoutsideIindex)The requirements for pure water in the pharmacopoeias of the United States, European Union, and Japan. |