Product Introduction
The new generation Milli-Q Reference system is capable of producing high-quality ultrapure water from pure water. After the first step of purification, the resistivity of pure water reached 18.2 M Ω. cm (at 25 ℃), and the TOC value was below 5 ppb. The produced water enters the Pak terminal purifier through a small circulation loop. Different terminal refiners can meet different application requirements.
main features
1. Stable and efficient production of ultrapure water
1) Pure water from Elix ™ Pure water system is preferred to enter Q-Gard according to the selected water source ™ The initial purification column and the pre purified water flow are then passed through a dual wavelength ultraviolet lamp to ensure that organic molecules are oxidized and bacteria are removed. Continuing with Quantum ™ The purification column removes ions and organic pollutants below trace levels to provide ultra pure water quality that meets application requirements. Ultra pure water meets the first level water quality standards set by ASTM, CAP, NCCL, EP, and USP.
2) Multiple water qualities can be combined with a series of terminal refiners to meet the needs of different specific applications
2. Comfortable and flexible water intake
1) Connecting a water intake arm with a 75 cm long water pipe makes water intake more convenient.
2) The quantitative function can automatically take the amount of water you need, saving you time. The ergonomically designed water intake bracket is suitable for laboratory glassware of various heights and shapes.
3) The main display screen of the Chinese interface can display detailed information about system operation and performance. Clear graphic display helps with specific operations such as maintenance.
4) Consumables can be easily replaced within 5 minutes, making maintenance easier.
5) The system will remind 15 days in advance to replace consumables or schedule service visits to ensure uninterrupted water supply.
6) The water intake arm can be upgraded to a POD water intake device, effectively saving laboratory space
Application scope
New Generation Milli-Q ® Reference system,Adhering to the consistent excellent quality, the performance is further improved to meet more diverse needsLaboratory requirements. This system produces high-quality ultrapure water from pure water, which can meet the needs ofVarious application requirements, and water quality meets or exceeds the various requirements to be followedRegulatory and standard requirements.
1. Selecting a 0.2 micron terminal filter to produce water is suitable for most physicochemical analysis applications such as HPLC, LC-MS, MALDI-TOF-MS, IC, ICP, AA, etc
2. BioPak water can be used for molecular biology applications and cell culture
3. Experimental study on EDS endocrine disruptors using EDS Pak produced water for environmental analysis
4. Choose VOC Pak to produce water that meets the requirement of completely non-volatile organic compounds
Technical Parameter
Grade I ultra pure water quality:
1. Electrical resistivity 18.2M Ω .cm@25 ℃
2 TOC ≤5ppb
3. Particles (>0.2 μ m/mL)<1(Install Millipak or BioPak as terminal refiners)
4. Bacteria<0.1 cfu/mL (with Millipak or BioPak installed as terminal refiners)
5. Heat source (endotoxin)<0.001Eu/mL (installing BioPak as terminal purifier)
6. RNases<1pg/mL (installing BioPak as the terminal purifier)
7. DNases<5pg/mL (Install BioPak as terminal purifier)