3S-CL Mn Manganese Ion (Standard Edition) Online Analyzer is an analytical instrument used for online analysis and process monitoring. It uses LED light source for colorimetric analysis and measurementmanganeseIonic (Standard Edition)ofConcentration.
The PAN photometric method is used to determine manganese. The PAN reagent is suspended in water by a surfactant and forms a chromogenic complex with manganese. The absorbance intensity is directly proportional to the manganese concentration in the sample, measured at 572 nm.
Product advantages
The instrument adopts a dual compartment housing to ensure complete separation between the electronic components (upper part) and the sampling and analysis components (lower part).
The instrument uses less reagent consumption, reducing operating costs.
The instrument can use an internal dilution module, with a measurement range ranging from trace μ g/L to 1000mg/L; Multiple chemicals can be supplied to meet actual needs.
The analyzer is equipped with a temperature heating module.
The analyzer can be equipped with a refrigerator to keep the chemicals at the appropriate temperature.
The equipment can be set up for automatic calibration, verification, and zeroing, which can reduce downtime and operator intervention, ensuring accurate results.
The analyzer adopts a color touch screen interface to display measurement values and status information. Easy to access menus and functional interfaces. And supports USB data download.
You can choose the dual channel version. By adding an external sequencer, the number of sampling channels can be increased to 4.
The instrument undergoes factory testing before leaving the factory. Simply connect the power supply, sample and reagent pipelines, and the analyzer can operate.
Product parameters
Measurement parameters |
Mn2+ (ppb,ppm,mg/L) |
size |
604*380*242mm (H*W*D) |
measuring principle |
Differential absorbance |
weight |
44 pounds (20 kilograms) |
measuring range |
1-200ppb Mn2+For a 26mm reaction tank, 5-1000ppb Mn2+For 16mm reaction tanks; Internal dilution up to 50ppm Mn2+ |
power supply |
110/220V AC 50/60Hz or 24 VDC Power consumption: MAX80 watts |
thoroughfare |
1,2 with integrated switch valve, 3,4 with external sequencer |
output |
Two 4-20mA analog signals; One Modbus RTU RS485 digital signal; |
Analyze frequency |
Freely programmable, batch close to continuous analysis |
call the police |
2 programmable SPDT passive relays |
cycle time |
8-10 minutes, including pre-treatment before the analysis cycle and cleaning after measurement. |
digital input |
Remote start/stop operation |
reactor |
Built in temperature heating module |
temperature |
41-113℉(5-45℃) |
water sample |
Non pressure overflow container Temperature: 41-122 ℉ (5-50 ℃) Flow rate: 80 to 500ml/minute Connection: 6Mm (¼ inch) |
humidity |
10% to 90% non condensing (indoor installation); Outdoor installation requires the use of protective cabinets or rainproof systems; |
drainage |
No pressure, standard atmospheric pressure drainage Connection: 12mm (1/2 inch) |
install |
Wall mounted (standard), Desktop support or panel installation (Option) |