1、 Product Overview | |
Metal rotor flowmeter is a commonly used variable area flow measurement instrument in industrial automation process control, which has the characteristics of small volume, large detection range, and easy use. It can be used to detect the flow rate of various liquids and gases, especially suitable for measuring the flow rate of low velocity and small flow media. The metal rotor flowmeter consists of a measuring tube, a float, and an indicator. The float moves freely up and down in the measuring tube, changing the flow area in the pipeline. As the flow rate changes, the vertical position of the float in the measuring tube also changes accordingly. The position of the float is accurately transmitted to the dial of the indicator through a magnetic transmission system, indicating the change in flow rate value. The LZ series metal rotor flow meter has both on-site indication and intelligent remote transmission types, with pointer display of instantaneous flow rate, LCD display of instantaneous and cumulative flow rate, standard 4-20mA current output, upper and lower limit alarm, HART protocol output and other functions, providing a wide selection space for different users. |
|
2、 Main features | |
● Reliable work, minimal maintenance, and long service life |
|
3、 Technical parameters | |
|
4、 Classification and Naming | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
LZ— Metal tube float flowmeter notes Z Basic type indicator indicator D Remote transmission indicator -15 DN15 nominal orifice -25 DN25 -40 DN40 -50 DN50 -80 DN80 -100 DN100 -150 DN150 / Down in, up out structure form H1 Left in and right out H2 Right in, left out AA Side in, side out LA Bottom in and side out S threaded M Quick connector connection R1 321stainless steel essence material R4 304stainless steel R6 316stainless steel R Hc4 blow one's breathCalloy Ti titanium F liningF46(PTFE) M7 Linear indicator indicator type M9 Nonlinear indicator (local pointer displays instantaneous flow) ESK 24VDCLong distance transmission4-20mAStandard current signal combination function ESK-Z 24VDCLong distance transmission4-20mAOn siteLCDDisplay instantaneous and cumulative flow ESK-D On site lithium battery power supply;LCDDisplay instantaneous and cumulative flow K1 Bring an online alarm point signal Bring down the online alarm point signal K12 Bring online and offline alarm point signals Ex Intrinsic safety explosion-proof explosion-proof type Exd EEx d L liquid Measurement medium Quality classification G gas T Measurement tube with insulation/insulating jacket increase function HT The temperature of the measuring medium is greater than120High temperature |
five.measuring range(Metal rotameter)
orifice (mm) |
Flow range |
Maximum pressure loss (kPa) |
accuracy |
||||
Water L/h |
Air m3/h |
||||||
Common |
Anticorrosive type |
Common |
Anticorrosive type |
water |
gas |
||
15 |
2.5~25 |
2.5~25 |
0.07~0.7 |
0.07~0.7 |
6.5 |
7.1 |
2.5 |
4.0~40 |
4.0~40 |
0.11~1.1 |
0.11~1.1 |
7.2 |
|||
6.3~63 |
6.3~63 |
0.18~1.8 |
18~1.8 |
6.6 |
7.3 |
||
10~100 |
10~100 |
0.28~2.8 |
0.28~2.8 |
7.5 |
|||
16~160 |
16~160 |
0.4~4.0 |
0.4~4.0 |
6.8 |
8.0 |
||
25~250 |
25~250 |
0.7~7.0 |
0.7~7.0 |
7.2 |
10.8 |
||
40~400 |
40~400 |
1.0~10 |
1.0~10 |
8.6 |
10.0 |
||
63~630 |
———— |
1.6~16 |
———— |
11.1 |
14.0 |
||
25 |
100~1000 |
63~630 |
, , , 3.0~30 |
1.6~16 |
7.0 |
7.7 |
|
160~1600 |
100~1000 |
4.5~45 |
3.0~30 |
8.0 |
8.8 |
||
250~2500 |
160~1600 |
7.0~70 |
4.5~45 |
10.8 |
12.0 |
||
400~4000 |
250~2500 |
11~110 |
7.0~70 |
15.8 |
19.0 |
||
50 |
630~6300 |
400~4000 |
18~180 |
11~110 |
8.1 |
8.6 |
|
1000~10000 |
630~6300 |
25~250 |
18~180 |
11.0 |
10.4 |
||
1600~16000 |
1000~10000 |
40~400 |
25~250 |
17.0 |
15.5 |
||
80 |
2500~25000 |
1600~16000 |
60~600 |
40~400 |
8.1 |
8.0 |
|
4000~40000 |
2500~25000 |
100~1000 |
60~600 |
9.5 |
13.0 |
||
100 |
6300~63000 |
4000~40000 |
200~2000 |
100~1000 |
10.0 |
11.0 |
|
10000~100000 |
———— |
300~3000 |
———— |
Note 1: Flowmeters with measurement ranges not listed in the table can be processed according to requirements
Note 2: Special calibration is at level 1.5
Note 3: Special specifications can be customized
6、 Connection method | |
![]() |
|
7、 Installation size | |
![]() |
![]() |
![]() |
|
![]() |
|
![]() |
|
![]() |
|
![]() |
|
8、 Alarm switch | |
Technical parameters and wiring of electric remote transmission signals Pre value alarm switch sensor The B1, B2, B3, B4 indicators can be equipped with 1 or 2 pre-set alarm switch sensors. |
|
![]() |
|
Two wire switches require connection to WE77 transistor amplifiers or equivalent products to output switch signals. The three wire switch requires a voltage supply of 10-30VDC, and the switch signal is output through a built-in signal amplifier for direct application with the PLC. |
|
9、 Electrical wiring diagram | |
![]() |
|
10、 Magnetic Filter Size Table | |
![]() |
![]() |