KSPY-9000 gas chromatograph
Instrument features:
1. The display window adopts 5.7The inch industrial color touch screen design provides more comprehensive display information and more reasonable interface operation;
2. Adopting 10/100MAdaptive Ethernet communication interface, built-inIPThe protocol stack is convenient for enterprises to realize long-distance data transmission through internal LAN and Internet; Facilitating laboratory setup, simplifying laboratory configuration and data management;
3Internal Design3An independent connection thread that can connect to local processing, unit supervisors (such as chief engineers, technical plant managers, etc.), and higher-level supervisory departments (such as environmental protection bureaus, technical supervision bureaus, etc.), facilitating real-time monitoring of instrument operation and analysis of data results by unit supervisors and higher-level supervisory units;
4. Equipped with NETChrom ®Workstation, capable of supporting multiple chromatography instruments(253Simultaneously work on the platform to achieve data processing and counter control;
5 NETChrom ®Built in workstationModbus/TCPServer, which can easily access the analysis resultsDCS(Distributed Control System);
6Adopting a modular structural design, the design is clear and easy to replace and upgrade, protecting the effectiveness of investment,Can meet the analysis of complex samples and can be equipped with various high-performance detectors, such asFID、TCD、ECD、FPDandNPDWait;
7. Using logarithmic amplifier design,Make the data results more accurate and reliable.
8Abandoning traditional pointer pressure gauges and being able to loadEPCtechnologyControl gasRoad control, automation level, and overall performance are close to first tier brands;
9Implemented self-protection against gas path faults, automatic ignition, shutdown, and automatic opening of the gas path, achieving one key start;
10. Design a timed self starting program that can easily complete online analysis of gas and liquid samples (with injection components required);
11The system is designed with an automatic sampler interface and comes with multiple built-in driver programs, allowing for the installation of an automatic sampler at any time;
Main technical indicators:
● Interface display: 5.7Inch industrial color touch screen display.
Temperature controlled area: 8Road; Support dual column box and dual rear door opening
Temperature control range: 4 ℃ to 450 ℃ above room temperature, increment: 1 ℃, accuracy: ± 0.01 ℃
● Program heating order: 16steps
●Cheng Sheng rate:0.1~39℃/min(Ordinary type);0.1~80℃/min(High speed type)
● External events:6Road; Auxiliary control output2road
● Injection methodFill column injection, capillary injection, six way valve gas injection, automatic headspace injection optional
● Number of detectors:3One;FID、TCD、ECD、FPDandNPDoptional
● Air path controlMechanical valve control methodEPCMethod selection
●EPC、EFCWorking mode:2Seed; Constant current mode, constant voltage mode
●EPC、EFCWorking gas:5Seed; Nitrogen, hydrogen, air, helium, argon
●EPC、EFCControl range: Pressure:0~0.6 MPa; Traffic0~100 mL/minperhaps0~500 mL/minair
●EPC、EFCControl accuracy: pressure0.01 KPa; Traffic0.01 mL/min
● Pressure sensor:
Accuracy: Full scale<±2%
Reproducibility:<±0.05 KPa
Temperature coefficient:<±0.01 KPa/℃
Range:0~0.6 MPa
● Flow sensor:
Accuracy: Full scale<±5%
Reproducibility:<±0.5%(full scale)
Range:0~500 mL/min
● Start the injection method: Choose between manual and automatic options
● Communication interface:Ethernet:IEEE802.3
Power supply: 220V ± 10%, 50Hz;2500W
● Volume: 572 × 552 × 465(High)mm
Weight: 50kg(About)
Technical specifications of detector
Hydrogen flame ionization detector(FID)
Detection limit: Mt ≤ 3 × 10-12 g/s(Hexadecane)-Isooctane solution
Noise: ≤ 5×10-14 A
Drift: ≤ 1×10-13 A/30min
● Linear range: ≥ ten6
Thermal conductivity detector(TCD)
Sensitivity: S≥3500 mV•ml/mg(Conventional)5000 mV•ml/mg(High sensitivity)(Benzene solution)(Zoom in)2、4、8Multiple Choice)
Noise: ≤ 10μV
Baseline drift: ≤ 30μV/30min
●Linear range:≥104
ECd(ECD)
●Detection limit:≤1×10-14 g/s
Noise: ≤ 0.03 mV
Baseline drift: ≤ 0.2 mV/30min
●Linear range:≥103
●Radioactive Source:63Ni
flame photometric detector(FPD)
●Detection limit:(S)≤5×10-11 g/s ,(P)≤1×10-12 g/s
Noise: ≤ 0.03 mV
Baseline drift: ≤ 0.2 mV/30min
●Linear range:≥103(S), ten2(P)