The KF-6901B ground wave ultrasonic comprehensive calibration instrument is mainly used to calibrate instruments and meters that use TEV (transient ground voltage) and ultrasonic sensing methods to detect partial discharge of high-voltage equipment in power. During testing, by detecting the output pole, the instrument will simulate the TEV signal and ultrasonic signal of the partial discharge characteristics of electrical equipment insulation faults, and the amplitude and frequency of the TEV signal can be set separately.
TEV calibration technical parameters
Output amplitude range: 0.02V-30V
Output frequency range: 20Hz-60KHz
Output polarity: Positive and negative polarity @ 0.02V-1V, positive and negative polarity @ 2V-30V
Amplitude accuracy: ± 0.5%
Frequency accuracy: ± 0.5%
Amplitude resolution: 2mV
Frequency resolution: 10Hz
Output step size: 1mV
Ultrasonic calibration technical parameters
Output range: 30-70dB
Output step size: 2dB
Output method: Open
Power supply: Built in rechargeable battery
Control software: calibrate software and communication interface through PC
Source mode: Two types of simulated pulse outputs: internal simulated partial discharge source and external standard pulse source