brand | Self operated brand | application area | Medical and health care, environmental protection, chemical industry, biotechnology, petroleum industry |
---|
install:
installJYQ-3000B typesolid samplerThe following steps should be followed:
⑴ Install according to the structural diagramsolid sampler.
⑵ Connect the power cord properly.
⑶ Adjust the height of the injector to ensure that the propulsion rod, injection boat, and cracking furnace are on the same horizontal line.
operate:
JYQ-3000B typesolid samplerIf installed properly, the following steps should be taken to put the system into use.
⑴ Place all power switches in the ON position.
⑵ Install the quartz boat and adjust the speed and progress of the sampler, utilizingPress the 'forward' and 'backward' buttons to adjust the position of the slider.
(3) Perform a leak test on the gas circuit before injection to ensure that there is no air leakage at the gas circuit joint
Note: When used in conjunction with a quartz boat, the slider should be able to fully insert the quartz boat into the furnace when moving forward, and it should be easy to position the quartz boat when moving backwardsolid samplerAt the injection port of the sample.
⑷ ensureinstrumentThe system has adjusted various parameters and conditions and conducted blank analysis on the quartz boat, waiting for sample analysis.
⑸ Place the prepared sample into the quartz boat.
according toinstrumentTime events in software systems enable automatic sample injection.Before conducting sample analysis, ensure that the quartz boat is sufficiently cooled.