Cermet II is a reliable intelligent instrument used for online humidity measurement of gases in many applications. The Cermet II includes an advanced ceramic sensor from Michell and a display control unit, which can be connected by a cable of up to 1200 meters. Cermet II is easy to use and suitable for almost all humidity measurement applications. It can display multiple measurement units, such as ° C or ° F dew point, ppmV, Lb/MMSCF or g/m ³, with a dew point range of -100~+20 ° C dp, a maximum pressure resistance of 30 MPa (300 barg), and allows dynamic pressure compensation through signal input from a pressure transmitter (provided as an optional option).
highlight
- Wide range
- ± 1.0 ° C accuracy
- Analog and digital outputs
- Pressure transmitter signal input
- Up to 4 channels of alarm relays
- Measurement units are optional
- Clear LED display
|
 |
technical indicators
|
| install thread |
5/8 'UNF (parallel) |
| Sensor body |
stainless steel |
| Sealing protection level |
IP54 (NEMA 12) with standard connectors, (IP66, NEMA 4X) with optional molded connectors |
| range |
Verification range -100~+20 ° C dew point |
| Operating Temperature |
-40~+60 ° C (compensation range -20~+40 ° C) |
| precision |
± 1 ° C-59~+20 ° C dew point; ± 2 ° C -100~-60 ° C dew point |
| resolution |
0.1 ° C-80~+20 ° C dew point; Dew point at 1.0 ° C-100~-80 ° C |
| instrument |
|
| unit of measurement |
° C, ° F dew point; ppmVRange and resolution are optional; lb/MMSCF, Range varies with pressure (natural gas); g/m³(Natural gas) |
| Auxiliary measurement parameters |
Pressure signal obtained through external sensor 0-30 Mpa Any user-defined industrial standard transmitter Automatic compensation conversion ppmV, lb/MMSCF, And g/m³Waiting for units |
| power supply |
85 ~ 265 V AC; 95 ~ 370 V DC |
| Sealing protection level |
IP54 (NEMA 4); IP66 (NEMA 4X) only refers to the front panel |
| Operating Temperature |
-20 ~ +50°C |
| relay |
Two 10 A/240 V relay alarms, user settings: alarm points can be set within the selected unit range, as well as sensor error alarms |
| output |
Analog quantity: 4-20 mA, corresponding range can be set and adjusted Digital quantity: RS232: Provides a bidirectional data transmission channel, allowing users to perform all parameter settings |
| Options: |
|
| power supply |
18 ~ 36 V AC; 9 ~ 60 V DC |
| relay |
Additional 5 A relays with 1 to 2 channels (up to a maximum of 4 channels in total), normally open type Alarm point setting+sensor error alarm |
| output |
Analog quantity: 0-10 V, corresponding range can be set and adjusted Digital quantity: RS485: Provides a bidirectional data transmission channel, allowing users to perform all parameter settings |
| cable |
Standard 2m (up to 1200m in length) |
|
|
|
|
|
|