PRODUCT DETAILS
oneOverview Describe
1.1 Purpose
Intelligent battery internal resistance testerBy adopting the most advanced AC discharge testing method, it is possible to accurately measure the voltage and internal resistance at both ends of the battery, and use this to determine the quality of the battery capacity and technical status. Customers can choose the internal resistance test of the battery according to their own situation、voltage testreachCapacity estimationThe result ofAs the basis for internal resistance matching when assembling new batteries; Testing the internal resistance of the battery before and after discharge is used to identify truly outdated batteries.
1.2 Characteristics
lIntelligent and digital, fully Chinese operation menu, accurate measurement, and simple operation.
lWeight not exceeding0.45KgHandheld design, single person operation, automatic measurement throughout the process.
lTo meet various battery internal resistance detection standards, a complete database of battery internal resistance parameters must be included, and the battery standard internal resistance can be defined according to different batteries.
lThe testing method is simple, will not affect the working state of the battery, and will not pose any safety hazards.
lThe instrument itself can store a large amount of test data and perform conclusive queries and analysis on the instrument. It can also use battery test data forUExport the disk to computer software to generate charts and curves for analysis.
lTest reports can be easily importedExcelandWordFile and print it as a report in the specified format for easy management and to reduce workload.
lFour terminal multi-purpose testing clip, integrating testing clip, probe and other functions, capable of adapting to98%The above battery connection installation methods and battery pole forms.
1.3 function
Serial Number |
project |
Description |
1 |
Battery measurement |
Internal resistance measurement and performance evaluation can be conducted for single cell batteries, group batteries, and battery+connection resistors (optional) |
2 |
data management |
Open playback, transfer to USB drive, delete, and format data records for data recording |
3 |
system management |
Including clock settings, parameter calibration, language selection, test waveforms, file management, and software version information |
1.4 Technical Parameters
project |
parameter |
measuring range |
Internal resistance: 0.0m Ω --120mΩVoltage: 0.000V--25v |
Connection resistance measurement range |
0.0mΩ—65.0mΩ |
Minimum measurement |
Internal resistance: 0.001 mΩ |
Resolution |
Voltage:1mV |
measurement accuracy |
Internal resistance:1%Voltage:0.15% |
display |
272×480 4.3Inch TFT LCD+touch screen |
Size |
186mm×98mm×40mm |
Weight |
0.45KG |
Memory capacity |
Store measurement data of 100000 batteries |
Working power supply |
4000mAHRechargeable lithium battery, fully charged and functional8~12hour |
External power supply |
AC100~240V/DC5V-1APower adapter/charger |
communication interface |
USB interface (can be plugged into a USB flash drive) |
2、 Operation guidance
Instrument startup/Shutdown
There is one on the left side of the instrument panelTurn the power switch to the upper end to turn it on, and turn it to the lower end to turn it off.Main startup interface:
The entire interface consists of three parts: top, middle, and bottom. On the upper left is the menu name, and on the right is the battery level display bar. The bottom edge is the time bar, and the middle is the display and operation area;
When using touch screen operation, you can directly click on the menu task item, and then click on the selected menu task item to execute it. If using keyboard operation, use“TAB”Press the key to switch the selected menu task item, and press confirm to execute.
2.2.1Customize standard parameters
clickThe "Unit" button on the right side of "Battery Voltage" will pop up a custom parameter dialog box. After setting the custom parameters, click "Save". After saving, you can see the button“V”Become 'customized'x”, indicating successful customization.
2.2.2Standard parameter selection
clickThe button on the right side of the "Battery Voltage" or "Standard Capacity" characters can pop up a list of standard parameters. Select a parameter according to your needs, click again to set the parameter, and the main interface will update and display. After clicking the "Save" button at the bottom of the "Group Measurement" interface,
The set parameters will be permanently saved, and will not be lost when turned on or off again. There is no need to set the parameters again, making it convenient for the next test.
2.2.3Operating instructions
clickThe "Operation Instructions" button displays a dialog box detailing the entire single section measurement operation process and precautions.
2.3 Data management function
Click on the main interfaceThe "Data Management" menu item enters the data management function interface, including single cell battery measurement data and group battery measurement data, which can be opened, replayed, and transferred for data recordingUDisk, delete and other operations, as well as formatting data records. The interface reference for data management is shown in the following figure:
2.4 Clock setting
When using a touch screen, you can directly select input items or buttons, and enter numbers by scrolling up and down. If operated using a keyboard,standard“TAB”Key switching input itemWhen selecting the year, month, day, hour, minute, and second, pressUse the "↑" and "↓" keys to increase or decrease numbers.
2.5 System Management
After entering the system settings, the system settings submenu will be displayed. Click on the task item to enter the system settings page, and click the return button to return to the main menu. The menu content is shown in the following figure:
2.5.1 File Management
The instrument can be used throughUTo upgrade the system software, first insert the upgraded software image file into itUDisk, shutdown, throughOTGConnect the wires properlyUDisk and device, press and hold the back button to power on, then release it, and wait for the upgrade. The interface will prompt the upgrade progress, and after the upgrade is successful, it will automatically start to the running interface.
2.5.2language selection
The software supports multiple languages. In the early stage, it first supports switching between Simplified Chinese and English languages, and the software needs to reserve a multi language switching mode. The language selection interface is shown below:
In the language selection interface, select the desired language, click confirm, and the display interface will switch. Press the back button to return to the system settings interface.
2.5.3test waveform
As an auxiliary function, it can measure0~200VThe DC voltage is convenient for users to conduct auxiliary tests.
2.5.4Version information
Double click on the system management menu pageThe "Version Information" menu item allows you to view the instrument system software version and product number, as shown in the figure:
3、 Instrument structure
3.1 Host
The host is used for on-site testing, operations, calculations, display of test results, storage, and other purposes.(Quantity: 1 unit).
3.2 Charger
The charger is used to power the host and charge the battery. When in use,Connect the charger input plug toOn a 220V power supply, the output terminal can be connected to the input terminal of the host power supply.quantity1).
3.3 User Manual
The user manual provides a detailed introduction to the usage functions and operating methods. Users should read carefully before using the instrument for measurement.quantity:1 volume).
3.4 ToolboxUsed to install the host and accessories. quantity1)
fourInstrument storage instructions
4.1 The FLASH chip used in the instrument stores data including startup screen data, measurement data, etc.
4.2 There are several types of instrument data:
1) Power on screen data: Logpic.BMP、 Length and width 272 * 480.
2) Test result data: File name: "AAACCCC. DAT", AAA is REM, CCCCC is serial number. Like REM00035.DAT
4.3 Operation of Instrument Data
In the main menu of the instrument panelUnder the "Data Management" function, data files can be processedopen、“export”(Transfer deposit)USB flash drive)delete(a separate piece of data),format(Delete all data) and other operations. Data file transfer toThe USB drive is in DAT file format, such as REM00035.DAT.