Huaxu CenturyHXMKThe series Coriolis mass flowmeter is designed based on the Coriolis principleMade by calculation, it is not affected by the physical characteristics of the measured fluid (such as temperature, pressure, viscosity, density) during the process of fluid flow detection, and can achieve direct and precise measurement of the mass flow rate of pipeline fluid.
Huaxu CenturyHXMKThe series Coriolis mass flowmeter is designed based on the Coriolis principleMade by calculation, it is not affected by the physical characteristics of the measured fluid (such as temperature, pressure, viscosity, density) during the process of fluid flow detection, and can achieve direct and precise measurement of the mass flow rate of pipeline fluid.
HXMKThe series flowmeter has no movable parts and no obstructionFlow detection design, wide range ratio, high accuracy (up to)±0.1%), withstand voltageFeaturing high temperature and wide temperature range, it can perform high-precision mass flow measurement and density/temperature measurement on various fluids such as oil products, natural gas, liquefied gas, compressed gas, solvents, slurries, and high viscosity fluids. It is widely used in various production and research occasions and is suitable for the majority of mass flow measurement industries.
Product Features:
● Multi parameter measurementInstantaneous/Cumulative/Mass/Volume Flow, Density, Temperature
● High precision (standard precision)0.2% reading, high-precision up to 0.1% reading)
● Measurement is not affected by the physical properties of the medium (temperature, pressure, viscosity, density)
● Suitable for media such as gases, liquids, and mixed media
● The maximum range ratio can reach100:1
● No obstruction design, sensor has no movable parts
● Optional sanitary type/Explosion proof type, suitable for various application scenarios
:Typical applications:
● Petroleum and Petrochemicals: Oil/Lipids/carbon black and other industries
● Special medium: high viscosity fluid/Gas liquid solid mixture medium
● Fuel industry: natural gas/Liquefied gas/compressed gas
● building materials/Glass/Paper Industry
● textile/Printing and dyeing/coating industry
● brewing/Food/Beverage/Pharmaceutical Industry
● sewage treatment/Transportation industry
● The Great Learning/Research Institute/Metrology Industry
:: Technical parameters::
Medium:Liquids, gases, and mixed media compatible with materials
Flow range:from0-5kg/h to 0-200t/h, please consult
Density range:0~3.2g/m , On site adjustable
Temperature range:-50℃~260℃
Liquid mass flow standard accuracy:±0.2% reading
High precision liquid mass flow rate:±0.1% reading
Gas mass flow accuracy:±0.5% reading
Density measurement accuracy:±0.005g/cc
Temperature measurement accuracy:±0.5℃
Repeatability:±0.1% reading
Zero stability:Based on the diameter of the pipeline, from0.002kg/h to 3kg/h
Zero point adjustment:On site adjustable
display:standard configurationLCD display screen, capable of displaying up to 3 decimal places
Dual row multi parameter display: Upstream display quality flow rate
Display the volume flow rate, temperature, density, and other cyclic parameters in the following rows
Standard voltage resistance:4MPa, Optional higher voltage resistance, please refer to the parameter table
Medium temperature:-50℃~260℃
Ambient temperature-20℃~70℃
Response time:0.1-5s adjustable (factory default 0.5s)
Output signal:Simultaneously possessing the following three output signals:
4-20mA (flow or density or temperature), active output
0-20kHz (flow only), compatible with 0-2kHz, 0-5
kHz、0-10kHz, Ordering requires specification
RS-485 (instantaneous/cumulative flow, density, temperature)
Route Two4-20mA output:Flow rate, density, or temperatureOn site adjustable
4-20mA output load capacity:750Ω
Pulse output electrical indicators:0 to+5VDC, 50% duty cycle
Liquid receiving material:Measuring tube:316L stainless steel (standard)
Process connector:304 stainless steel (standard), 316L optional
Shell material:304 stainless steel (standard)For other liquid and shell materials, please consult for customization
Process connection:Please refer to the parameter table for optional light pipes, card sleeves, and flanges
Light tube:1/4 inch card sleeve: 1/4 inch
Flange:ASME/ANSL CLASS 150、CLASS 300
Class 600, depending on pressure and temperature
Power supply:DC 24V ± 10% overall power consumption: 8W
Protection level:IP67
Order Information:
● Flow range: minimum flow, normal flow, maximum flow
Installation method: integrated or split type
● Medium pressure:
● Medium temperature:
Explosion proof situation:
Output signal: 4-20mA output is flow rate, density, or temperature
● Medium information: medium name, pH, viscosity, density, gas-liquid solid state
Frequency output signal: Within the range of 0-20kHz, a specified output frequency value is required
Measurement tube material: The standard material is 316L stainless steel, and other materials can be customized upon consultation
● Process connector material: Standard is 304 stainless steel, 316L is optional, others need to be consulted
● Process connection method: light tube size, or card sleeve size, or flange specification
Other accessories or requirements: such as high precision, second 4-20mA output, HART protocol, etc