Specific parameters:
model |
PBV-012 |
PBV-025 |
PBV-040 |
PBV-070 |
PBV-100 |
|
Chamber volume(L) |
125 |
250 |
400 |
700 |
1000 |
|
controller |
Programmable, touch screen controller(3.5"Color Display Screen |
|||||
temperature range(℃) |
-25 ~100 |
|||||
temperature variation 1) 2) exist25℃time(±℃) |
0.3 |
0.3 |
0.3 |
0.4 |
0.2 |
|
Temperature fluctuations2) (±℃/℉) |
exist-25℃time |
0.3 |
0.3 |
0.4 |
0.4 |
0.4 |
exist-15℃time |
0.3 |
0.3 |
0.4 |
0.4 |
0.4 |
|
exist-10℃time |
0.3 |
0.3 |
0.4 |
0.4 |
0.4 |
|
exist25℃time |
0.3 |
0.3 |
0.4 |
0.4 |
0.4 |
|
exist70℃time |
0.4 |
0.4 |
0.6 |
0.6 |
0.6 |
|
exist100℃time |
0.6 |
0.6 |
0.7 |
0.8 |
0.8 |
|
heating time2) (minute), follow25arrive100℃ |
34 |
34 |
52 |
35 |
38 |
|
Cooling time2) (minute), follow25arrive-25℃ |
44 |
44 |
34 |
33 |
42 |
|
When the temperature recoversbetween2)After the door is opened(minute) |
exist-25℃time |
5.0 |
5.0 |
5.0 |
4.0 |
7.0 |
exist-15℃time |
4.5 |
4.5 |
3.5 |
5.5 |
5.0 |
|
exist-10℃time |
6.0 |
6.0 |
4.5 |
4.0 |
5.0 |
|
exist25℃time |
5.0 |
5.0 |
4.0 |
5.5 |
3.5 |
|
exist70℃time |
5.0 |
5.0 |
9.5 |
5.0 |
7.5 |
|
exist100℃time |
5.5 |
5.5 |
9.0 |
6.0 |
7.0 |
|
Programming control |
temperature range(℃) |
-12.5 ~ 87.5 |
||||
Heating rate(℃/divide) follow-12.5℃arrive87.5℃ |
2.6 |
2.6 |
1.5 |
2.7 |
2.0 |
|
refrigeration rate(℃/divide) follow87.5℃~-12.5℃ |
2.1 |
2.1 |
2.8 |
2.5 |
2.4 |
|
Number of steel wire framesmeasure(standard configuration/Top up) |
2 /13 |
2 /18 |
2 / 23 |
2 / 30 |
2 / 33 |
|
Shelf roomdistance(mm) |
30 |
|||||
Each shelf carries a weightmeasure(kg) |
15 |
20 |
30 |
40 |
45 |
|
All load-bearing capacitymeasure(kg) |
50 |
70 |
90 |
120 |
150 |
|
size(W*D*H) |
Internal dimensions(mm) |
500×500×500 |
600×650×650 |
750×700×800 |
900×8001000 |
1000×9101100 |
Dimensions(mm) |
1060×702×1340 |
1154×852×1540 |
1304×902×1724 |
1454×1540×1628 |
1720×1650×1534 |
|
net weight(Kg) |
160 |
260 |
320 |
390 |
440 |
|
communication interface |
standard configurationRS-232interface(Optional)RS-485Interface) |
|||||
Power requirements(230V, 50Hz) |
1ph /15.5A |
1ph / 25.2A |
3ph /17.4A |
3ph / 26.7A |
3ph / 29A |
|
Item number |
AAHK5012K |
AAHK5022K |
AAHK50312K |
AAHK50412K |
AAHK50512K |
|
Power requirements(380V, 50Hz) |
3ph /10A |
3ph /15.4A |
3ph /16.8A |
|||
Item number |
- |
- |
AAHK5038K |
AAHK5048K |
AAHK5058K |
1)conform toIEC60068-3-5requirement.
2)conform toDIN 12880requirement.
※ The above parameter values are230V/50Hz,380V/50HzMeasured under certain circumstances.
※Due to continuous parameter updates, please understand if there has been no notification.
※ doFDARegistered company under the item, product inFDARegister.
P - Series
Performance characteristics
It can have special purposes, such as heat treatment and drying treatment.
Microfabrication PID controller/automatic calibration function/automatic adjustment function.
Heat resistant silicone can provide a strong sealing effect, preventing heat loss inside the cavity.
A 2-node latch can also improve sealing performance.
-
A good insulation system can have better energy utilization efficiency.
-Even if the system operates at a high maximum temperature, the surface temperature of the equipment will not exceed 51 ℃ (in accordance with DIN EN 563 standard)
-
The door has a reliable design.
-When the door is opened 90 degrees, it can carry a weight of 50kg at a distance of 30cm from the door latch
Two point door latch+stainless steel chamber |
Punched shelves can carry heavy objects (optional) |
Independent over temperature limiter |
safety features
After power outage and restoration, the operator can choose to operate automatically or manually.
Over temperature protection.
Current overload protection and leakage protection (earth leakage circuit breaker)
-
Independent safety fuse
-Important parts such as controllers and heaters have short-circuit protection. Especially each heater has an independent fuse, which can prevent all short circuits.
-
Reverse phase protection.
-The electrical connection error indicator can help users quickly repair connection problems.
-
Current overload protection and leakage protection
-(Ground leakage circuit breaker)
-
Door opening alarm function.
-When the door is open, the system will automatically turn off the heater and fan.
-
Wire damage alarm function.
-When the surface of the wire is damaged, it will automatically shut down.
-
Built in door lock and standard key.
Alarm signal light (optional)
Simple operation
The inner cavity is made of corrosion-resistant stainless steel.
-
Efficient utilization of space and convenient placement of samples through internal cavity design.
-Comes with two steel wire frames as standard, capable of carrying heavy samples.
-Optional perforated shelf for carrying overweight (40kg) (optional)
-
A simple and easy-to-use LCD controller.
-The control signal of the device can be promptly fed back through the color LCD touch screen.
-The set temperature value and the current temperature value can be displayed on the screen simultaneously.
-Support programming functionality.
-Displaying the operation process in graphical mode makes the data easier to understand.
-The automatic mode can be turned on or off, and can be programmed to run automatically.
-The monitor can display abnormal temperature changes or power failures.
-The interface supports three languages (English/Chinese/Korean)
-6-channel digital recorder can record temperature and other parameters (optional)
-
Computer communication interface
-The RS-485 communication interface (optional) can connect 32 units. Multiple units can be monitored and operated simultaneously.
-With this software, programming will become even simpler.
-Can check the running status in real-time. The data is displayed in graphics, making it easier to understand.
-It has storage and programming functions.
-Data can be saved in EXCEL format or document format.
-It can store program information, set value backups, reset, and temperature values.
-Standard RS-232 communication interface.
describe |
No |
Number of programming groups |
100 |
Number of repetitions per group |
999 |
Number of repetitions per paragraph |
255 |
Number of segments per group |
100 |
Number of segments used* |
6000 |
Each programming period |
0 ~ 99Hour 59minute |
*Although each group has 100 programming segments, a maximum of 6000 segments can be used instead of 10000 segments. Otherwise, the programming mode needs to be reset
Color LCD controller |
Can be connected to a computer for remote control and data collection |
Temperature can be programmed for automatic operation and controlmake |