HI2300 is a microcomputer controlled, conductivity/total solid solubility/sodium chloride desktop measuring instrument. Its unique automatic range selection function can automatically identify and select the range based on the measured sample, without the need to replace the probe. The instrument adopts manual or automatic temperature compensation, and users can choose the temperature coefficient value independently. This function can also be turned off for actual conductivity measurement. The instrument also has a stability indicator to remind users when to record measurement results. HI2300 also has GLP performance, and the stored data can be input into the computer through the RS232 interface. In addition, users need to enter a unique ID number to identify the instrument.
Technical specifications of HI2300 desktop conductivity meter/HI2300 pure water conductivity meter
Measurement indicators | Conductivity (EC) | Total Solid Solubility (TDS) |
measuring range (Temperature compensation at 20 ℃ or 25 ℃) |
0.00 to 29.99μS/cm or 30.0 to 299.9μS/cm 300 to 2999μS/cm or 3.00 to 29.99mS/cm 30.0 to 200.0mS/cm Up to 500.0mS/cm (measure conductivity value without temperature compensation) |
0.00 to 14.99ppm or 15.0 to 149.9ppm 150 to 1499ppm or 1.50 to 14.99g/l 15.0 to 100.0g/l Up to 400.0g/l (TDS value measured without temperature compensation) |
Resolution | 0.01μS/cm(0.00 to 29.99μS/cm) 0.1μS/cm(30.0 to 299.9μS/cm) 1μS/cm(300 to 2999μS/cm) 0.01ms /cm(3.00 to 29.99ms/cm) 0.1mS/cm (above 30.0mS/cm) |
0.01ppm(0.00 to 14.99ppm) 0.1ppm(15.0 to 149.9ppm) 1ppm(150 to 1499ppm) 0.01g/l(1.50 to 14.99g/l) 0.1g/l (above 15.0g/l) |
accuracy | ± 1% reading ± (0.05 μ S/cm measurement value) | ± 1% reading ± (0.03ppm measurement value) |
Correction point | Equipped with six calibration points for automatic recognition | Conductivity correction and TDS automatic recognition correction simultaneously |
Conversion factor | __ | 0.40 to 0.80 (factory default value of 0.50) |
temperature compensation | Automatic or manual compensation from 0 to 60 ℃ | Automatic or manual compensation from 0 to 60 ℃ |
Temperature compensation constant | 0.00 to 6.00%/℃ (factory default value 1.90%/℃) | |
Measurement indicators | Salinity (NaCl) | Temperature (℃) |
measuring range | 0.0 to 400.0% | 0.0 to 60℃ |
Resolution | 0.10% | 0.1℃ |
Accuracy | ± 1% reading | ±0.4℃ |
Correction point | Single point automatic recognition calibration (recommended calibration solution HI7037) | 2 temperature calibration points at 0 ℃ and 50 ℃ (adjustment interval ± 1 ℃) |
Supporting electrode | HI76310 has ATC function, four ring high-performance conductive electrode (conductivity constant K=1) | |
data interface | RS232 | |
Supporting power supply |
HI710006 dedicated 220V/12VDC power supply |
|
application environment | 0.0 to 50.0 ℃ 95% RH | |
Size/Weight | 240×182×74mm / 1200g |
Standard configuration of HI2300 desktop conductivity meter
HI2300 | Measurement host, HI76310 with ATC four loop conductive electrodes and related accessories |
HI2300P | Measurement host, HI76310 with ATC four loop conductive electrodes, carrying case, and related accessories |
HI2300 desktop conductivity meter selection accessories
HI76310 | ATC high-performance four ring conductive electrode |
HI710006 | Dedicated 220V/12VDC power supply |
HI76404 | Electrode specific bracket |
HI920010 | 9-9 data connection cable |
HI92000 | Specialized software package for data processing and upgrading |
HI2604 | Moisture proof, waterproof, and shockproof carrying case |