We have launched the latest 12mm fermentation fluorescence dissolved oxygen sensor for the fermentation industry, which adopts internationally leading fluorescence lifetime technology, RS-485 output, and supports MODBUS protocol. Compared with traditional polarographic oxygen sensors, it has many technical advantages such as no preheating, no electrolyte, no oxygen consumption, no generation, high accuracy, anti-interference, maintenance free, fast response, and more environmentally friendly. Acid and alkali resistant, capable of high-temperature disinfection, in compliance with CIP disinfection standards and steam disinfection standards.
Measurement principle
The fluorescence method dissolved oxygen analyzer is based on the quenching principle of specific substances on active fluorescence in physics. Blue light emitted from a light-emitting diode (LED) illuminates the fluorescent material on the inner surface of the fluorescent cap. The fluorescent material on the inner surface is excited and emits red light. By detecting the phase difference between the red and blue light and comparing it with the internal calibration value, the concentration of oxygen molecules is calculated, and the final value is automatically compensated by temperature and pressure.
Unique Highlights
Suitable for steam disinfection, high-temperature sterilization, and CIP cleaning
No film, no electrolyte required, no polarization
It does not consume oxygen and is not affected by flow rate
Built in temperature sensor, automatic temperature compensation
Small drift, fast response, and more accurate measurement
Free from maintenance, long applicable cycle
Comparative test results of optical dissolved oxygen analyzer in citric acid fermentation tank
TECHNICAL INDEX
Type
|
Parameter values
|
Dissolved oxygen range
|
0-20mg/Lperhaps0-200%saturation
|
reaction time
|
10 sec
|
Pressure Range
|
0-4 bar
|
Dissolved oxygen accuracy
|
±1%
|
sensor drift
|
<1% per year
|
Maximum disinfection temperature
|
130℃
|
operation temperature
|
0-50℃
|
Temperature Accuracy
|
±0.2℃
|
sensor interface
|
supportRS-485, MODBUSagreement
|
25℃drift
|
<1% /circumference
|
dress join in marriage
|
PG 13.5thread
|
power information
|
DC 5.5V +/-10%, current<50mA
|
temperature sensor
|
NTC
|
Shell material
|
stainless steel
|
Shell diameter
|
12 mm
|
Shell length
|
199.3 mm
|
school accurate
|
One or two point calibration
|
ordering information
model
|
describe
|
H503-B
|
12mmFermentation specific fluorescent dissolved oxygen sensor
|
H610-A
|
Online transmitter, one digital input; two-way4-20mAOutput; One digital output
|
Optional Accessories
model
|
describe
|
H9004
|
12mmFermentation dissolved oxygen fluorescent cap
|