▏Product Features
◎This machine adopts a 5.7-inch LCD vertical display with a large Chinese screen, providing clear and intuitive images. 10 temperature control values (column box 1, 2, injector) can be displayed simultaneously on the same interface
1, 2, 3, temperature set values and actual values for detectors 1, 2, 3, auxiliary 1, 2, etc., and reserved areas for fault alarm positions, time display areas, event areas, etc.
◎It can display the temperature, time, events, range rise, stopwatch, extreme temperature protection values of the instrument, and add status indicator lights, fault alarm lights, etc. on a universal basis
Integrating culture and convenience.
◎Having core temperature control development technology, PID temperature control system, utilizing the characteristics of bus development technology, adopting control chip, dual over temperature protection of temperature control software and hardware, enabling temperature control
The system runs for a long time, is reliable and durable
▏Technical parameters
◎Pressure protection method: The carrier gas has pressure protection function. When the gas pressure is below 0.1mpa, the instrument automatically cuts off the power supply, effectively protecting the safety of chromatographic column use
◎Display mode: 320x240 Chinese vertical LCD display
◎Temperature control area: 7-way temperature control
◎Temperature control accuracy: ± 0.1 ℃ - ± 0.2 ℃
◎Temperature control range: room temperature ± 8 ℃ -450 ℃, with increments of 1 ℃
◎Program heating: 16 stage program heating, heating rate 0.1-40 ℃/min, increments of 0.1 ℃, initial and termination time range: 0-6000min, increments of 1min
◎Extreme temperature: maximum protection value 450 ℃
◎外部事件:8通道外部事件
◎Power supply: 220V 50HZ 2000W
▏Detector indicators
Hydrogen flame ionization detector (FID):
Quality based detector
Suitable for trace analysis of organic compounds
◎Ignition method: automatic ignition
◎Detection limit: Mt ≤ 5 × 10-12g/s (n-Hexadecane-Isooctane Solution)
◎Noise: ≤ × 10-14A
◎Drift: ≤ 1 × 10-13A/30min
◎Linear range: ≥ 106
Thermal conductivity detector (TCD):
Concentration type detector
Constant analysis applicable to organic and inorganic gases
◎Sensitivity: S ≥ 3500mV • ml/mg (conventional)5000mv • ml/mg (highly sensitive) (benzene toluene solution)
◎Noise: ≤ 0.01mv
◎Baseline drift: ≤ 0.03mV/30min
◎Linear range: ≥ 104
Electron Capture Detector (ECD):
Radioactive ionization detector
Suitable for trace analysis of compounds containing heteroatoms such as N, O, and S
◎放射性同位素: 63NiActivity:370MBq(10mCi)
◎Detection limit: ≤ 1 × 10-14g/s
◎Noise: ≤ 0.03mv
◎Baseline drift: ≤ 0.2mv/30min
◎Linear range: ≥ 103
Flame photometric detector (FPD):
Quality based detector
Suitable for trace analysis of sulfur-containing and phosphorus containing compounds
◎Detection limit: (S) ≤ 5 × 10-11g/s ,(P)≤1×10-12g/s
◎Noise: ≤ 0.03mv
◎Baseline drift: ≤ 0.2mv/30min
◎Linear range: ≥ 103(S),102(P)
Nitrogen phosphorus detector (NPD):
Quality based detector
Suitable for trace analysis of nitrogen-containing and phosphorus containing compounds
◎Detection limit: (N) ≤ 5 × 10-12g/s ,(P)≤1×10-12g/s
◎Noise: ≤ 0.03mv
◎Baseline drift: ≤ 0.2mv/30min
◎Linear range: ≥ 103(N),102(P)
