WM-6800 Fully Automatic High Temperature and High Shear Viscosity Tester
Product Features:
1. Implement standards SH/T0703 and ASTM D5481. It can be widely used in departments such as petroleum, chemical, metallurgy, power, transportation, commodity inspection, and scientific research.
2. The core host adopts TI's AM3354 processor, Cortex-A8 core, and a 1GHz clock speed; The operating system adopts Windows Embedded Compact 7 real-time industrial control system. Thoroughly abandoning the use of nuclear free and non operational microcontrollers, the modernization of instrument control has been truly achieved, ushering instruments into a new era of intelligence.
3. The monitor adopts the original 7.0 inch 800 × 480 pixel true color TFT-LCD LCD screen from Innolux, with a fully Chinese operation interface and delicate, intuitive and generous display.
4. The operation adopts a capacitive touch screen, which has the advantages of no mechanical loss, dust and water resistance, resistance to radio frequency interference, and long service life.
5. Historical data storage adopts FLASH data storage, which can store 2010 historical data. The data can be saved for 10 years without loss, and the stored data cannot be changed.
6. Rear USB interface, which can be used for later software updates and can also be connected to a mouse for precise operation.
7. The temperature detection adopts the original imported PT100 platinum resistance temperature sensor, combined with a high-precision AD converter, to control the heating of the electric heating tube through excellent linear mathematical models and unique control algorithms, making temperature measurement and control faster, more accurate, and stable.
8. The automatic high-temperature and high shear viscosity tester is the first of its kind in China, with key components made of CN accessories from the United States to ensure product quality;
9. The automatic high-temperature and high shear viscosity tester adopts a one click start function to automatically detect the results and display the data on the screen, making the detection data more intuitive and controllable;
10. The automatic high-temperature and high shear viscosity tester uses specialized high shear viscosity calculation software, which prompts operation and is easy to learn;
11. There is a plug valve on each sample tube of the automatic high-temperature and high shear viscosity tester, which meets the standard requirements;
12. The automatic high-temperature and high shear viscosity tester is equipped with FTC circuit protection device, which automatically cuts off the power supply when overheating occurs.
Technical Indicators:
1. Heating method: electric heating rod heating
2. Temperature control range: normal temperature~150 ℃
3. Control pressure range -350kpa to 3500kpa
4. Sample capacity -8ml
5. Test unit: 1 unit
6. Capillary viscometer size: ¢ 0.15 × 16mm
7. Pressure display: Touch screen display
8. Time display: Touch screen display
9. Result Calculation: Specialized Calculation Software
10. Working power supply: AC220V 50HZ