Product Features:
1. The unique membrane treatment technology and accessories in the United States can adapt to different municipal tap water in China;
2. The unique source water purification pretreatment Wtr Res Clean system can extend the service life of consumables and ensure the long-term normal operation of the instrument;
3. The reverse osmosis system has automatic, manual, and timed flushing functions, effectively protecting and extending the service life of the RO membrane;
4. Bilingual control switching between Chinese and English (default to Chinese at the factory), more in line with your language habits;
5. The multifunctional FLOM control system has alarm prompts for no water, consumable failure, consumable expiration, water leakage, and full water, automatic fault detection function, water quality parameter setting exceeding alarm function, as well as storage function for various operating data and customer password protection function, ensuring the security of operating data;
6. Unique three pump design and fully enclosed pipeline design for regular automatic circulation during non water intake periods to maintain stable water quality;
The 7.5.1 inch LCD screen can intuitively display various operating parameters, including water quality and temperature parameters of the inlet, RO water, and UP water channels;
8. It can simultaneously produce pure water and ultrapure water, with automatic continuous water production and adjustable operating parameters;
9. The water production system and electrical system are completely separated, completely eliminating safety hazards;
10. The enterprise has obtained EU CE certification and ISO9001-2008 international quality management system certification;
11. The equipment complies with the national laboratory water standard GB/T6682-2008, the Chinese national standard for electronic grade ultrapure water (GB/T11446.1-1997), and the American pharmaceutical grade water standards (CAP, ASTM, NCCLS);
Optional:
1. Water leakage alarm module, beep alarm, can also receive FLOM module to send SMS alarm to the user's mobile phone (SIM card required by the user);
2. Wireless control module, capable of remotely accessing RO/UP water;
3. According to customer requirements, it can be changed to three water outlets;
4. SMS control/alarm module: provides water leakage alarm, remote control, remote parameter modification, and information group sending functions;
5. R232/USB communication interface, can be connected to a printer to print water quality parameters;
Application fields:
Semiconductor rinsing water, optical lenses, high-purity quartz, crystal sand, nanoorganisms
Central water supply systems for materials, ultrasonic cleaning, steel enterprises, universities, and research institutions.
Technical Specifications:
|
Product Name |
FYY Basic Application Type |
FFX Basic Analysis Type |
FBZ standard reagent type |
FCR heat removal type |
FDY low organic type |
FJY Gene Research |
|
Produced water quality |
Third grade water |
Secondary water |
First grade water |
First grade water |
First grade water |
First grade water |
|
Product Model |
FYY(30-120)01/02 |
FFX(30-120)01/02 |
FBZ(30-120)01/02 |
FCR(30-120)01/02 |
FDY(30-120)01/02 |
FJY(30-120)01/02 |
|
working conditions |
Inlet: Municipal self source water pressure: 0.15 Mpa-0.5 Mpa Inlet temperature: 5-40 ℃ |
|||||
|
Conductivity (μ S/cm) |
≤5 |
≤0.5 |
≤0.067 |
≤0.055 |
≤0.055 |
≤0.055 |
|
Electrical resistivity (M Ω. cm) |
≥0.2 |
≥2 |
≥15 |
≥18.25 |
≥18.25 |
≥18.25 |
|
TOC content (ppb) |
- |
- |
≤10 |
≤5 |
≤2 |
≤2 |
|
Endotoxin content (EU/ml) |
- |
- |
- |
≤0.001 |
- |
≤0.001 |
|
Particulate matter (0.22 μ m) |
- |
- |
≤1/ml |
≤1/ml |
≤1/ml |
≤1/ml |
|
Bacteria (CFU/ml) |
- |
- |
≤1 |
≤1 |
≤1 |
≤1 |
|
Water intake flow rate (L/min) |
1-2 |
1-2 |
1-2 |
1-2 |
1-2 |
1-2 |
|
Water production flow rate (L/h) |
30-120 |
30-120 |
30-120 |
30-120 |
30-120 |
30-120 |
|
Metal cations (ppb) |
- |
- |
Fe、Al、Cr、Ni、Cu<0.005ppb Zn、K<0.02ppb、Na<0.01ppb |
|||
|
Anion (ppb) |
- |
- |
Cl-、SO4 2- <0.01ppb、NO2-、NO3-<0.02ppb |
|||
|
Soluble Silicon (SiO2)/(mg/L) |
- |
<0.02 |
<0.01 |
|||
|
Absorbance (254nm. 1cm optical path) |
- |
<0.01 |
<0.001 |
|||
|
Dimensions (mm) |
600 × 540 × 1350 (length x width x height) mm |
|||||
|
weight |
80kg |
|||||
|
Model Analysis FBZ (30-120) 01/02 Standard Reagent Type (30-120) refers to a water production rate of (30-120L)/H (30-120) 01 refers to single-stage reverse osmosis, and (30-120) 02 refers to two-stage reverse osmosis |
||||||