Product features:
1. The unique membrane treatment technology and accessories in the United States can be adapted to different municipal tap water in various regions;
2. The reverse osmosis component is automatically flushed upon startup, and after flushing is completed, it automatically switches to the water production state, effectively protecting and extending the service life of the RO membrane;
3. Low pressure protection for water inlet, automatic shutdown when the water tank is full;
4. Built in high-voltage protection device and automatic pressure relief function;
5. Simultaneously produce two types of water quality: pure water and ultrapure water; Meet your various experimental requirements;
6. Real time display of water temperature monitoring shows the current production water temperature measurement range: 0-45 degrees Celsius
7. Intelligent control system: multiple status displays, with clear visibility of the operation status of water free, full, water production, and flushing equipment;
8. All pipelines, joints, and water storage containers used in the system are NSF (National Health Foundation) certified to ensure their sanitary quality;
9. Configure a 3.2G vacuum pressure water tank;
10. The equipment complies with the national laboratory water standard GB/T6682-2008, the Chinese national standard for electronic grade ultrapure water (GB/T11446.1-1997), and the American pharmaceutical grade water standards (CAP, ASTM, NCCLS);
11. The enterprise has obtained EU CE certification and ISO9001-2008 international quality management system certification;
Application areas:
The classic series of ultrapure water machines can completely replace traditional distilled water purifiers that are time-consuming, energy consuming, and inefficient, and can also be used to provide higher pre-treatment for producing high-quality ultrapure water.
Technical parameters:
product name
FYC Basic Application Type
FFC Basic Analysis Type
FBC standard reagent type
Produced water quality
Third grade water
Secondary water
First grade water
PRODUCT MODEL
FYY(5-15)01/02
FFX(5-15)01/02
FBZ(5-15)01/02
working conditions
Inlet: Municipal self source water pressure: 0.15 Mpa-0.5 Mpa Inlet temperature: 5-40 ℃
Conductivity (μ S/cm)
≤5
≤0.5
≤0.067
Electrical resistivity (M Ω. cm)
≥0.2
≥2
≥15
TOC content (ppb)
-
-
≤10
Endotoxin content (EU/ml)
-
-
-
Particulate matter (0.22 μ m)
-
-
≤1/ml
Bacteria (CFU/ml)
-
-
≤1
Water intake flow rate (L/min)
1-2
1-2
1-2
Water production flow rate (L/h)
5-15
5-15
5-15
Soluble Silicon (SiO2)/(mg/L)
-
<0.02
<0.01
Absorbance (254nm. 1cm optical path)
-
<0.01
<0.001
Dimensions (mm)
400 × 405 × 570 (length x width x height) mm
weight
20kg
Model Analysis FBC (5-15) 01/02 Standard Reagent Type
(5-15) refers to water production rates of 5L, 10L, and 15L/H. (5-15) 01 refers to single-stage reverse osmosis, and (5-15) 02 refers to two-stage reverse osmosis