Yancheng Galaxy Technology Co., Ltd
Home>Products>FCG-5H Explosion proof Air Sampler (Explosion proof Individual Dust Sampler)
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
    13851330428
  • Address
    3rd Floor, Building 2, Wuzhou International Trade City, Mingzhu Road, Jianhu County
Contact Now
FCG-5H Explosion proof Air Sampler (Explosion proof Individual Dust Sampler)
FCG-5H Explosion proof Air Sampler (Individual Dust Sampler)
Product details

FCG-5Htype explosion-proofemptyGas samplerIndividual dust sampler

1、 Overview

FCG-5H typeAir samplerIt is based on GB50325-2010, GB/T 17061-1997, HJ/T 400-2007, JJG 956-2013《Calibration of atmospheric samplerDeveloped in accordance with the basic requirements for air samplers such as the "Regulations" and the "GBZ/T 192 Determination of Dust in Workplace Air" standard, it is used for long-term or individual sampling of total dust, respiratory dust, and metal dust using (1-5) L/min flow membrane sampling clamps. Optional flow range: (1000~5000) mL/min

The FCG-5H air sampler consists of a pump, control circuit, power supply, and other components. Can run continuously under normal circumstances12More than hours, convenient for sampling work. It has real-time flow display, sampling volume display, sampling time setting, sampling site temperature, atmospheric pressure measurement display, power indicator, button set flow, no need to adjust knob valves. Its compact structure, small size, light weight, easy to carry, sturdy and durable, strong load capacity, accurate and stable flow display, low noise, etc. It can be carried around during work and is convenient for users to useConstant current air sampler.

2、 Instrument composition:

The sampler consists of a self-developed vacuum pump (with core technology and low maintenance costs), a microcomputer controller, a multifunctional LCD display, a sealed battery pack with repeated charging, discharging, overcurrent and short-circuit protection, and a digital flow component. The casing of this series of instruments is made of polycarbonate injection molding, which has flame retardant, moisture-proof, and anti-static properties.

3、 Main features:

1The entire working process of the sampler is completed by microcomputer software and hardware, which is reliable, easy to operate, and highly intelligent;

2The sampler is equipped with a self-developed vacuum pump (with core technology and low maintenance costs in the later stage). Users can choose between explosion-proof or non explosion-proof types (with identical mechanical characteristics). The pump has a small volume, low long-term heat generation, reliable operation, low power consumption, and high vacuum degree;

3Using an electronic flowmeter, the flow rate value can be displayed in real-time on the LCD screen and the sampling volume can be accumulated;

4The instrument uses explosion-proof lithium battery packs (imported lithium batteries), which have no battery memory effect, large capacity, long lifespan, and are equipped with our company's dedicated charger for good charging effect and safety;

5Adopting a short-circuit current protection device to ensure that the sampler operates within an inherently safe current range, giving the sampler good explosion-proof performance;

6There are two sampling methods for this instrument: workers working in the workplace can hang it on their belts for long-term sampling, and a matching tripod can be placed on the work site for repeated short-term sampling.

fourTechnical parameters

lFlow range:(1000~5000)mL/min conform toClass B (4.5kPa)Flow regulation step size500mL/min

Can withstand 5Kpa resistance @ 5000mL/minFlow regulation step size500mL/min

Equipped with individual respiratory dust sampling head, individual total dust sampling head, sampling belt, dust measuring filter membrane, etc.

lConstant current modeElectronic constant current,Built in precision electronic flow meter, users can calibrate flow rate with different loads. Ensure the authenticity and reliable quality of each sampling flow.

lFlow error:≤±5%

lTraffic stability:≤5%

lTraffic repeatability:≤2%

lTiming range:1min~99h59min.

lCan be setThree timing modes

l 1-Cyclic sampling, recyclable99Next, set the sampling time, pause time, and number of cycles

l 2-Manual sampling: After setting the sampling flow rate, you can directly sample, time up, and manually shut down

l 3-Timed sampling, set the sampling time, and stop sampling when the time is up.

lTiming accuracy:≤±0.1%

lWorking temperature:-10℃~50℃

lContinuous working hours:12hour

lExplosion proof form: intrinsic safety typeEx ib BT3Gb

lHost dimensions:110mm×105mm×55mm

lWeight:≤0.4Kg

Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!