High performance spectrophotometric determination, continuous wavelength of 320-1100nm, wavelength accuracy: ± 1nm; wavelength reproducibility: ± 1nm, combined with dual beam optical characteristics in the visible light range, making measurement more convenient and flexible. It can measure nearly 150 water quality indicators such as COD, TOC, residual chlorine, total chlorine, total phosphorus, total nitrogen, ammonia nitrogen, etc. [Click to view all measurement items];
It can create user-defined programs, with dual modes of factory calibration and user-defined calibration. It has a large color display screen, user-friendly interface, 5000 sets of data storage, PLC-3 printer interface, USB and Ethernet interfaces, which can be used for data management and equipment updates. It is suitable for rapid sample analysis and measurement in laboratories and on-site, and widely used in scientific research, metrology, sample testing, and identification.
The chemical oxygen demand (COD) is determined using the potassium dichromate method (K2CrO7). The pre made COD reagent complies with the Standard method 5220D and US EPA 410.4 standards. The dedicated heating digestion device is made of heat-resistant and corrosion-resistant materials, and can digest 24 samples simultaneously. It has overheating protection function to ensure safe and stable use
Performance Features Highlights
Microcomputer control technology, countdown function, user-friendly interface, 320-1100nm continuous wavelength, automatic wavelength selection;
Adopting unique SMD microcomputer control technology, self calibrating tungsten lamp light source, holographic grating, silicon photodetector, 4 nm bandwidth; Absorption range: -3.3 to 3.3 Abs, backlight display screen.
The application with over 150 built-in water quality testing methods can be identified through barcodes and colorimetric cup types;
▷ Multi functional measuring cell, can use 13mm, 16mm, 24mm circular colorimetric dishes and 10mm, 20mm, 50mm rectangular colorimetric dishes, with good GLP function and automatic shutdown setting.
▷ 5000 sets of data storage, equipped with PLC-3 printer interface, USB and Ethernet interface, can be used for data management and device updates, convenient for data organization and recording, and easy to connect with computers for data transmission.
xd7x00_mb4_04.jpg
Technical Specifications
ET99735
measurement items
Built in [Click to view all measurement items] nearly 150 measurement parameter applications
optical system
Self calibrating tungsten lamp light source, holographic grating, grating monochromator, 4 nm bandwidth; Absorption range: -3.3 to 3.3 Abs
Wavelength range of 320 to 1100 nm (continuous wavelength); Wavelength accuracy: ± 1nm; wavelength reproducibility: ± 1nm
Measurement pool specifications
Multi functional measuring pool, capable of using 13mm, 16mm, 24mm circular colorimetric dishes and 10mm, 20mm, 50mm rectangular colorimetric dishes
GLP function with built-in real-time clock, background light display, and splash proof operation
data management
5000 sets of data storage, equipped with PLC-3 printer interface, USB and Ethernet interfaces; Convenient for data transfer with PC
Corrosion resistant operation panel, touch keyboard with key prompt sound
Language Selection
German, Chinese, English, Spanish, French, Italian, Portuguese, Japanese and other multi country interface selection
Power mode
AC230V/12VDC power adapter [recommended ET71310020]
Applicable Environment
5 to 40 ° C, RH: 30 to 90% non condensing
Dimensions & Weight
Host size: 275 × 150 × 270mm, host weight: 3.2kg
certification standard
CE Certification
Technical specifications of heating and digestion processor
Dissolve quantity
Can simultaneously digest 24 samples, with a digestion aperture of 16mm, suitable for international custom tube reagents
Temperature setting
70, 100, 120, 150 ° C four gear selection
time control
30, 60, 120 minutes or continuous heating. After digestion is complete, it will automatically turn off with a prompt sound
temperature range
Pt100A built-in temperature probe, accuracy: ± 0.3 ° C
Protection type
ABS heat-resistant and corrosion-resistant material, temperature automatic control, anti overheating system protection
Power mode
230 V/50-60 Hz or 115 V/50-60 Hz; 40W
Manually switch between 110 V and 220 V through the voltage switch on the back
Applicable Environment
10 to 35°C、RH:85% Max
Dimensions & Weight
Host size: 422 × 195 × 323mm, host weight: 4.5Kg
certification standard
EC Directive 2014/35/EC EN 61010-1:2010
Recommend purchasing accessories
ET99955 customized specialized chemical oxygen demand [COD-LR] reagent, applicable range: 0 to 150 mg/L (ppm)
ET99974 customized specialized chemical oxygen demand [COD-MR] reagent, applicable range: 0 to 1500 mg/L (ppm)
ET99106 customized chemical oxygen demand [COD-MR] reagent, applicable range: 0 to 15000 mg/L (ppm)
ET2420803 customized special COD standard solution, standard value: 100mg/L, specification: 30mL
ET2420804 customized special COD standard solution, standard value: 500mg/L, specification: 30mL
ET2420805 customized special COD standard solution, standard value: 5000mg/L, specification: 10mL
ET71310020 custom dedicated power adapter, specification: AC230V/12VDC
ET197635 customized glass colorimetric dish cleaning cloth, packaging: 4 pieces/group
ET418957 customized ∅ 16mm ten hole colorimetric dish test tube rack, suitable for heating and cooling of tube reagents