The ET3302A fully automatic nitrogen blowing concentrator provides a safe, effective, and automatically controlled sample concentration method while saving laboratory space.
Product Features
● Save laboratory space without the need for fume hoods
The patented endpoint sensor does not require manual monitoring
There are two specifications for glass components: 50 and 200ml
● Convenient and easy-to-use display and diagnostic functions
The temperature of the temperature controlled water bath can range from room temperature to 99 ℃
Automated operation can process 6 samples simultaneously
The airflow can be adjusted from 4-25psi
The features of the standard automatic fault diagnosis function include automatic diagnosis and discrimination of various faults such as endpoint sensors, heater faults, solenoid valve faults, etc
FEATURES
Number of samples | 6 50ml or 200ml sample tubes |
Maximum sample volume | 200ml |
evaporation rate | The evaporation rate in most applications is 2-8ml/min, depending on the temperature of the water bath and the gas blowing speed |
Terminal volume | Optional. The automatic endpoint sensing point should be 0.5ml, 1.0ml, or blow dried. You can also set the endpoint freely with a manual or timed controller |
water bath temperature | The temperature can be adjusted up to 95 ℃. Optical sensors should be removed in working environments above 60 ℃ to avoid damage |
Gas blowing requirements | The gas source should have a minimum pressure of 35psi and be equipped with a pressure regulating device. The gas inlet of ET3302 is a 1/4 inch tube. The typical working pressure level is 10-18psi. |
Gas exhaust port and pipeline | Randomly provide 5.1cm inner diameter: 3.8m long emptying pipe |