DTD16 Urine Iodine Determination Dissolver
DTD-16 Urine Iodine Determination Digestion Instrument User Manual 1 Overview This instrument won the first prize of the 1995 Heilongjiang Provincial
Product details
User Manual for DTD-16 Urine Iodine Determination Dissolver
1 OverviewThis instrument won the first prize of the 1995 Heilongjiang Provincial Health Department Science and Technology Lecture Award and the second prize of the Heilongjiang Provincial Science and Technology Progress Award. This instrument is suitable for sample digestion in metal element determination and is also suitable forSample digestion for the determination of non-metallic elements iodine, selenium, and arsenic. This provides reliable support for more analytical instruments such as atomic absorption spectrometers, polarometers, and flow injection analyzers to play a greater roleQuick sample processing methods. This product is the first of its kind in China. It is a new energy-saving product. Widely used in fields such as food analysis, environmental sample analysis, and biological sample analysis.
Figure 1 DTD -16 Constant Temperature Digestion Instrument
Performance and improvements of the 2DD-16 constant temperature digestion instrument:
The thickness of the bottom and surrounding areas of the constant temperature block has been increased, the heat capacity has been expanded, and the temperature stability has been improved.
The appearance size (mm) is 430 X 330 X165. The casing is made of stainless steel plate.
When approaching the set temperature, the automatic use of gradient voltage reduction to power the heating wire prolongs the lifespan of the instrument.
Heating hole arrangement: 60 digestion holes; The four corners are areas of poor temperature conduction, so there are no digestion holes.
3 Performance indicators:Temperature control range: Room temperature * 150 ℃; Temperature stability is better than 2 ℃; Digital display temperature. Power consumption: 800W.
4 Usage:Usage and placement environment: Place the DTD-16 digital constant temperature digestion instrument in a ventilated kitchen for use. Insert the power plug into a 220V AC power supply. After use, turn off the power and remove the DTD-16 digital constant temperature digestion instrument from the ventilated kitchen. Urine iodine testing must use digestive tubes from the same batch, with an outer diameter of 15 millimeters and a length of 100 millimeters. D15 × L100 urine iodine determination specialized digestive tube, can be reused repeatedly.
Set the digestion temperature: As shown in Figure 1, press the SET key, and the PV window will display SP. Press the ∧ and ∨ keys, and the SV window will display the adjusted temperature. Press the SET key to end the temperature setting. In the heating state, as shown in Figure 2, the PV window displays the current temperature. After 15 minutes, the set temperature can be reached and heated at a constant temperature.
List 5: Host, power cord (1 * * 220V AC), instruction manual, certificate of conformity.
,