Conventional five parameter monitoring instrument for water qualityCPT3200-AQ5s
Application:
Groundwater, wastewater, pond aquaculture water, soil solution, industrial wastewater, etc;
No need to replace the probe, the integrated structure of the sensor can synchronously measure 5-8 water quality parameters at once;
Flexible application mode, convenient testing can be carried out in both laboratory and field;
-
Continuous on-site monitoring is possible;
Performance advantages:
No need to replace the probe, the integrated structure of the sensor can synchronously measure 5-8 water quality parameters at once;
No need for any additives; Water quality indicators can be directly measured; Multi parameter probe with stainless steel waterproof casing.
Display: 240128 LCD display;
Power supply: AC220V or DC24V optional
-
Signal output: RS485 output, Modbus RTU protocol output. Analog signal output, requires additional output module (optional)
Conventional five parameter monitoring instrument for water qualityCPT3200-AQ5s
working principle:
CPA8-AQ5S/8SMulti parameter online water quality analyzerThe analyzer uses the principles of ion selective electrodes and light scattering electrodes for analysis work.
Combine electrode probes with 5-8 parameters.
The sensor unit includes a sensor probe and a stainless steel housing, and the measurement program has operation and data storage functions. Users can independently complete the operation and no longer need to send samples for testing elsewhere, saving time and money.
-
All parameters can be measured simultaneously, and since no additional reagents are used, this measurement system is fast and easy to operate, eliminating the need for laboratory analysis skills.
CPT3200-AQ5s
-
Electrode technical parameters:
Detection parameters
Technical Parameter
conductivityTDS
Principle: Electrochemistry
electrodeconstant:1.0
measurerange:0~2000μs/cm
Resistance press:6MP
Material:316Ltitanium alloy
dissolved oxygenDO
measuring range: 0~20.0 ppm
Accuracy: 2.0%,perhaps+/- 0.05%FS
Usage temperature: 0~50℃
Pressure: <1.7 to 3 kgf/cm2.
PH
measuring range: 1.0~14.0
Temperature:0~50℃;
Accuracy: +/- 2.5%R;
Temperature: 0~50/100℃;
Drift: <1%R /day;
reaction time:Da Da90% <1s~2min;
Pressure:Solid electrodes can>1.7kgf/cm2
turbidityTurb4210
Principle:Turb4220Forward scattering and transmission turbidity; Hengguang Strong Dual Probe Measurement
measuring range: 0~1000NTU
Accuracy: 2~5% FS
repeatability: <1%R
divide distinguish rate: 0.01NTU;
Usage temperature: 0~50℃;
Pressure: <10 kgf/cm2.
temperature
Measurement range:0℃~50℃
CPT3200-AQ5s
Beijing Beidouxing Industrial Chemistry Research Institute
Beijing Beidouxing Technology Co., Ltd
: