characteristic
According to medical standards, the dry heat sterilization operation at 160 ℃ for 2 hours is equipped with a program that can maintain a constant temperature of 160 ℃ for 2 hours.
● Dry heat sterilization can be carried out without removing the CO2 sensor. With high-precision and long-life CO2 sensors, even without removing the CO2 sensor, dry heat sterilization at 160 ℃ can be performed, and previous concerns about CO2 sensor contamination have been eliminated.
The pollution minimization inner box, shed board, and humidification plate are all made of antibacterial stainless steel, taking pollution prevention measures to the extreme.

Specifications
model |
IP610 |
|
mode |
Gas sleeve type with dry heat sterilization function |
|
Performance |
Temperature control range |
room temperature+5~50℃(37℃ operation requires room temperature to be within32Below ℃ |
Temperature regulation accuracy |
±0.1℃(at 37℃) |
|
Temperature distribution accuracy |
±0.25℃(at 37℃) |
|
Use humidity range |
95%±5%RHabove |
|
CO2Concentration adjustment range |
0~20% |
|
CO2Concentration adjustment accuracy |
±0.2%(at37℃CO2Concentration setting5%) |
|
Composition |
touch screen |
7Inch color touch screen |
Exterior Decoration/door |
Chemical resistance coating on the surface of electrical galvanized steel sheet |
|
Inner groove/Inner door |
Stainless steel plate/Tempered glass |
|
Door opening and closing method |
Standard right opening door (can be changed to left opening door) |
|
Thermal insulation material |
glass fibre |
|
Temperature inside the tank/Sink water temperature sensor |
Pt100Thermal resistance |
|
CO2sensor |
Infrared mode (new single light, two wavelength mode, no driver) |
|
Independent over lift prevention sensor |
KType thermocouple |
|
Sink heater |
130W×2 |
|
door/Door frame heater |
98W/25W(Output can be set arbitrarily) |
|
Lead hole |
Φ32mmback |
|
other |
Adjust the feetCO2Sample mouth(OUT)Temperature output terminalCO2Concentration output terminal, external alarm output terminal, various record displays, trend chartsUSBData export port, drain valve (Φ)10.5) |
|
Safety device |
Overcurrent leakage protection switch, independent over rise protector, abnormal temperature upper and lower limitsCO2Abnormal upper and lower limits of concentration, self diagnostic function, door opening alarm, humidification water shortage alarm (thermistor), button lock |
|
Specifications |
External dimensions |
W600×D664(765)×H880mm |
Internal dimensions |
W485×D540×H640mm |
|
Internal volume |
about167L |
|
Shed board size/Number of layers/spacing |
W471×D540/12layer/40mm |
|
Shed board material/bearing |
Antibacterial shed board (made of stainless steel)/about7kg/layer |
|
Power Supply(50/60Hz)Rated current |
AC220V 2A |
|
Accessories |
Shed board4Silicone plugs are used for parts, humidification plates (stainless steel), and lead holes2OneCO2Gas supply hose2m, hose clamp |