1、 Product Introduction:

1. Temperature range: -40 ℃ to 180 ℃

2 APT.lineTMPreheating chamber technology

3. Programmable condensation protection for samples

4. Heated observation window with LED internal lighting

5. BINDER communication software APT-COMTM 3 Basic version

6. Fault diagnosis system using optical and sound alarms

7. Controller using interval programming

8. Access hole with silicone plug 50mm, left

9. 4 casters, 2 with brakes

10. Computer interface: Ethernet

11. 230 V socket on the right side operating component

12. Display through LCD color monitor

13. Adjustable slope function function function

14. Integrated continuous automatic recorder

15. Real time clock

16. Door heating

17. 1 socket made of stainless steel

18. BINDER testing certificate

19. Internal chamber made of stainless steel

20. PT100 temperature sensor

21. Heat insulation through polyurethane foam

22. Fluorine free refrigerant R-404A

23. Cooling through a compression cooler

24. Finned evaporator

2、 Technical parameters:

Product model MK115-400V1 MK115-400V-C1
Choose a model standard Equipped with transformer and frequency converter
Item code 9020-0175 9020-0290
measure
Internal volume [L] 115 115
Net weight of equipment (unloaded) [kg] 260 260
Maximum total load [kg] 60 60
Maximum load per partition [kg] 30 30
The dimensions of the casing do not include additional components and connectors
Net weight width [mm] 980 980
Net weight at height [mm] 1725 1725
Depth net weight [mm] 865 865
Wall to side distance [mm] 300 300
Wall distance from rear [mm] 210 210
Observation window width [mm] 288 288
Observation window height [mm] 222 222
Internal dimensions
Width [mm] 600 600
Height [mm] 480 480
Depth [mm] 400 400
The number of doors
Outer door 1 1
Data related to the environment
Sound pressure level [dB (A)] 62 67
firmware
Number of partitions (hours/maximum) 1/4 1/4
Temperature performance data
Temperature range [℃] -40…180 -40…180
The temperature deviation depends on the set value [± K] 0.1…2 0.1…2
The temperature fluctuation depends on the set value [± K] 0.1…0.5 0.1…0.5
The average heating rate is based on IEC 60068-3-5 [K/min] 5.3 5.3
Cooling time from 180 ℃ to -40 ℃ [min] 90 90
The average cooling rate is based on IEC 60068-3-5 [K/min] 5 5
The maximum thermal compensation is at 25 ℃ [W] 2000 2000
Electrical data
Rated voltage [V] 400 480
Power frequency [Hz] 50 60
Rated power [kW] 3.5 3.5
Equipment safety device [A] 16 16
Phase (rated voltage) 3~ 3~

1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. The technical data assumes a fan speed of 100%. We reserve the right to make technical changes.