BINDER KB240 Low Temperature Incubator
1、 Product Introduction:
1. Temperature range: -5 ℃ to 100 ℃
2 APT.line ™ Internal preheating technology
3. Compared to the previous generation product, it can save up to 30% energy consumption
4. Cooling through a compression cooler
5. Door heating
6. Adjustable fan speed
7. Controller using interval programming and real-time programming
8. Interior door made of safety glass (ESG)
9. Two sockets made of stainless steel
10. 4 casters, 2 with brakes
11. Access hole with silicone plug 30mm, left
12. Independent adjustable temperature safety device, Class 3.1 (DIN 12880), equipped with optical and sound alarms
13. Computer interface: Ethernet
2、 Technical parameters:
Product model | KB240-230V1 | KB240UL-120V1 |
Choose a model | standard | standard |
Item code | 9020-0202 | 9020-0304 |
measure | ||
Internal volume [L] | 247 | 247 |
Net weight of equipment (unloaded) [kg] | 171 | 171 |
Maximum total load [kg] | 100 | 100 |
Maximum load per partition [kg] | 30 | 30 |
The dimensions of the casing do not include additional components and connectors | ||
Net weight width [mm] | 925 | 925 |
Net weight at height [mm] | 1465 | 1465 |
Depth net weight [mm] | 800 | 800 |
Wall distance from rear [mm] | 100 | 100 |
Wall to side distance [mm] | 100 | 100 |
Internal dimensions | ||
Width [mm] | 650 | 650 |
Height [mm] | 785 | 785 |
Depth [mm] | 485 | 485 |
The number of doors | ||
Inner door | 1 | 1 |
Outer door | 1 | 1 |
Data related to the environment | ||
Energy consumption at 37 ℃ [Wh/h] | 270 | 270 |
Sound pressure level [dB (A)] | 53 | 53 |
firmware | ||
Number of partitions (hours/maximum) | 2/9 | 2/9 |
Temperature performance data | ||
Temperature range [℃] | -5…100 | -5…100 |
The temperature deviation is 37 ℃ [± K] | 0.2 | 0.2 |
Temperature fluctuation at 37 ℃ [± K] | 0.1 | 0.1 |
The recovery time after opening the door in 30 seconds is 37 ℃ [min] | 2 | 2 |
Electrical data | ||
Rated voltage [V] | 200…240 | 100…120 |
Power frequency [Hz] | 50 | 60 |
Rated power [kW] | 1.2 | 1.2 |
Equipment safety device [A] | 16 | 16 |
Phase (rated voltage) | 1~ | 1~ |
1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. The technical data assumes a fan speed of 100%. We reserve the right to make technical changes.