Brand Model
Product Name:Airflow salt spray test chamber
Product model: KB-Y-60
Product Introduction
Airflow salt spray test chamberAfter surface treatment of various materials, including coatings, electroplating, inorganic and organic films, anodizing, anti rust oil and other anti-corrosion treatments, the corrosion resistance of the products is tested.
Control method
1.KB-Y-60Airflow typeThe salt spray test chamber adoptsThe Bernoulli principle absorbs salt water and atomizes it uniformly, without any obstruction or crystallization, ensuring the standard of continuous testing.
2. Nozzle: made of special glass nozzle, which can adjust the size of spray volume and spray angle.
3. spray volume: spray tower is set in the middle of the laboratory to make the fog fall evenly, and can be adjusted to 1~2 ml 80 cm2/hr (16 hour average volume)
Technical Specifications
|
studio dimensions |
600×400×450mm (L×W×H) |
|
analysisprecision |
± 0.1℃ |
|
controlprecision |
± 1.0℃ |
|
distributionessencedegree |
± 2.0℃ |
|
Test Method |
1.Salt water spray test: NSS, ACSS
(1) Laboratory: 35℃±1℃
(2) Saturated air tank: 47℃±1℃
2.Acetate salt spray test:NSS,ACSS
(1)Laboratory:50℃±1℃
(2)Saturated air tank:63℃±1℃
Add acetic acid with ice to adjust the pH value of the test solution to 3.1~3.3
3. Copper speed test: CASS
(1) Laboratory: 58 ℃±1℃
(2) Saturated air tank: 65 ℃±1℃
Add acetic acid with ice to adjust the pH value of the test solution to 3.1~3.3, and mix with 0.02~0.025% copper chloride |
product structure
|
laboratory |
adoptP.V.CPolyvinyl chloride board, thickness5Mm, heat-resistant up to 80℃ |
|
sealing cover |
Made of acrylic impact resistant board, thick6Mm, transparent sealing cover, real-time observation of the condition of the tested product |
|
pressure tank |
adoptSUS#304Stainless steel high-pressure resistant bucket, with the best insulation effect |
|
heating system |
Using imported stainless steel finned heat pipe U-Type electric heater to heat the air |
|
shelf |
Using a planar indexing frame, the angle can be adjusted arbitrarily, and the four sides are completely consistent in terms of falling and receiving fog, with a large number of test pieces placed. |
|
Control Panel |
◆ Laboratory temperature controller0-99.9℃, LCD digital display hasPIDAutomatic calculation, control error ±0.3℃
◆ Saturated air tank temperature controller0-99.9℃, LCD digital display hasPIDAutomatic calculation, control error ±0.3℃
◆ Heating tank liquid expansion safety temperature controller0-120℃
◆ Saturated air tank liquid expansion safety temperature controller0-120℃
◆ Digital display time controller0.1s-9990hr
◆ Accumulator0-9990hr
◆ Swing board illuminated buckle switch, capable of continuous buckle pressing20000time |
protective device
1Automatic power-off device when the water level is low
2Automatic power-off device when overheating occurs
3Equipped with safety warning light device
main controller
Communication contactor: Taiwan
Thermal relay: Taiwan
Communication relay: OMRON,
Time Relay: Taiwan
Controller: Taiwan
Overtemperature protector: Taiwan
Main power switch: Taiwan
Solid state relay: Taiwan
Button switch: Taiwan
Dry burn protector: Italy
Additional notes
| Voltage used: |
1.power supply:AC1¢3W 220V 50/60HZ※ Voltage fluctuation allowable value:Freeze voltage ±10% 2. Maximum load capacity:3.0KVA 3. Maximum operating current:15A |
| use water want Request: | ROReverse osmosis pure water or distilled water |
| Usage environment: |
Allow use in an environment of 0~30 ℃ Performance guarantee environment: 5~25 ℃ The installation ground should not have obvious inclination No other items can be placed within 60CM around the instrument |
| Accessories: |
1.LType storage rack 4 pieces 2Standard measuring cylinder 1branch 3Temperature indicator needle 2branch 4.OType storage rack 6branch 5. Collector 80㎝ 1branch 6Glass nozzle 1branch(Please replace regularly) 7. Sodium chloride test drug(NACL)500gram 2 bottles 8. One 2.5m long low resistance rubber cable for the power cord |
| Items excluded from valuation: |
1. Preparation project for setting up the venue 2. Input side electrical equipment 3. Natural drainage piping engineering 4. Drainage outlet equipment 5. Gas source 6. Mist exhaust port |
| Provide documents: |
1. Provide a copy of the order specification sheet when placing an order 2. Provide a copy of the user manual upon delivery 3. Provide a copy of the product warranty card upon delivery |
Airflow typeThe salt spray test chamber is one of the essential equipment for conducting salt spray tests on products. Nowadays, there are many manufacturers who make salt spray test chambers,But the quality variesNot even,When choosing salt sprayWhen testing the test box, it is recommended that you look for enterprises with high brand awareness, long establishment years, and non-standard customization capabilities, where product quality is guaranteedObstacles, and relatively good after-sales serviceRelatively complete.