ATDS-10AType fully automatic thermal desorption instrument(Second time)
1Instrument Introduction
ATDS-10AThe fully automatic thermal desorption instrument isBeijing Zhongyi Yusheng Technology Co., Ltd. has developed and launched a fully automated design, touch screen display, more convenient operation, and continuous operationA new generation fully automatic thermal desorption instrument for 10 samples.
ATDS-10A Fully Automatic Thermal Desorption InstrumentCan be used with all imported and domestic gas chromatographs, cansatisfyThe following standards:
1.《HJ/644-2013 Determination of volatile organic compounds in ambient air - Adsorption tube sampling - Thermal desorption gas chromatography-mass spectrometry method ";
2.《HJ/T400-2007 Sampling and Determination Method for Volatile Organic Compounds and Aldehydes and Ketones in Vehicles ";
3.《GB/T18883-2002 Indoor Air Quality Standards;
4.《HJ/583-2010 Determination of Benzene Compounds in Environmental Air Solid Adsorption/Thermal Desorption Gas Chromatography;
5. GB/50325-202Indoor environmental pollution control in civil construction projectsstandard》;
6《HJ734-2014 Determination of volatile organic compounds in fixed source exhaust gas Solid phase adsorption thermal desorption/gas chromatography-mass spectrometry method》Wait, wait.
2、 InstrumentCharacteristics and main functions
1. It can run automatically-10 samples, no need for personnel on duty;
2. Power on self-test, fault alarm and prompt, automatic positioning and calibration of sample disk;
3. Microcomputer program control, with main functions including:
⑴ Method parameter setting, real-time animation display of working status, and running time;
⑵ The desorption zone, injection valve, sample transfer tube, and secondary desorption zone are all separately heated and temperature controlled;
⑶ Set up the analysis program and press the run button to automatically complete the analysis of all samples;
⑷ The GC and chromatographic data processing workstations can be started synchronously, and the device can also be started by external programs;
4. The machine comes with the function of simulating and sampling standard samples, which makes it more convenient to create working curves through a thermal desorption instrument;
5. The built-in sampling tube activation program can be loaded to automatically activate the desorbed sampling tube;
6. Through time programming, automatic functions such as desorption, blow adsorption, re desorption, injection, and backwash cleaning can be achieved;
7. Adopting electronic refrigeration and second-order thermal desorption process to ensure narrow chromatographic peak shapes;
8. The sample transfer tube and injection valve have automatic blowback function, which avoids cross contamination of different samples;
9. For active substance analysis, an elastic quartz tube can be selected as the sample transfer tube;
10. The injection needle is easy to replace and can be connected to all models at home and abroadGC injection port.
3、 InstrumentMain technologiesparameter
1Desorption 1Temperature control range:
room temperature—400℃ In increments1℃Appointment;
2.Temperature control range of valve injection system:
room temperature—220℃ Increase by 1℃Appointment;
3.Temperature control range of sample transfer pipeline:
room temperature—240℃ In increments1℃Ren Shi
(For operational safety, the temperature control of the transmission pipeline adopts low-voltage power supply);
4. Desorption 2Temperature control range:
room temperature—400℃ In increments1℃Appointment;
(Heating rate)2000℃/min)
5. Temperature control range of cold trap:
-35℃—50℃In increments1℃Ren Shi
(Using electronic refrigeration device, no need for liquid nitrogen refrigeration)
6.Temperature control accuracy:±0.1℃ ;
7Temperature control gradient: ±0.1℃;
8.Sample grade:10 people
9Desorption recovery rate:>98%(Related to components);
10.Backblowing cleaningTraffic:0~600ml/Min (continuously adjustable);
11. Simulated sampling flow rate: 100ml/min;
12.Instrument size:450×360×510mm3;
13.Instrument weight: approximately30kg。