Used Shimadzu gas chromatograph, Shimadzu 2010 gas chromatograph, Shimadzu 2014 gas chromatograph
Used Shimadzu gas chromatograph, Shimadzu 2010 gas chromatograph, Shimadzu 2014 gas chromatograph GC-2010AF host FID detector; Split/No Split Injectio
Product details

Used Shimadzu gas chromatograph, Shimadzu 2010 gas chromatograph, Shimadzu 2014 gas chromatograph



Shimadzu GC-2010 Reference Technical Indicators

1. Column temperature box:

  • Operating temperature: room temperature+4 ℃~450 ℃ (when liquefying CO2:- 50℃ ~ 450℃);

  • Temperature setting: 1 ℃ ratio;

  • Temperature deviation: within 2 ℃;

  • Correlation of room temperature changes: 0.10 ℃/℃;

  • Number of temperature program segments: 20 segments;

  • Program ratio setting range:- 250 ~ 250℃/min;

  • Total program time:~ 9999.99min;

  • Heating rate: 50~300 ℃, about 3 minutes;

  • Cooling rate: 450~50 ℃, about 6 minutes;

2. Electronic carrier gas flow/pressure control:

  • Range of carrier gas flow rate setting: 0~1200ml/min (number of program segments: 7 segments);

  • Range of carrier gas pressure setting: 0~970Kpa (number of program segments: 7 segments);

  • Fixed line velocity type: can maintain a constant average line velocity of the chromatographic column during the heating process of the column temperature chamber;

  • *An additional 14 APC channels can be used to control the flow of detector gases such as tail gas, hydrogen gas, air, and diaphragm cleaning gas;

  • AFC can digitally set column pressure, total flow rate, diversion ratio, column flow rate, and column flow velocity;

3. Injection port:

Can simultaneously load 3 injection ports, each equipped with electronic carrier gas control

(1) Split/non split capillary column injection port:

  • *Operating temperature: 450 ℃;

  • Pressure setting range: 0~970Kpa;

  • Total flow setting range: 0~1200ml/min;

  • Setting range of diversion ratio: 0 to 9999.9;

  • Suitable for capillary columns with a diameter of 0.1mm~0.53mm;

(2) Wide caliber capillary column/packed column injection port:

  • *Operating temperature: 450 ℃,

  • Pressure setting range: 0~970Kpa;

  • Total flow setting range: 0~1200ml/min;

  • Suitable for packed columns and 0.53mm capillary columns;

4. Detector:

*Multiple detectors can be loaded simultaneously, with each detector equipped with electronic gas control

(1) Hydrogen flame ionization detector (FID):

  • *Operating temperature: 450 ℃;

  • Automatic ignition;

  • *Small inspection measurement:< 3pgC/s;

  • Dynamic range: 107

(2) Electron Capture Detector (ECD):

  • *Operating temperature: 350 ℃;

  • *Small inspection measurement:< 8fg/s (benzene trichloride);

  • Dynamic range: 104

(3) Flame photometric detector (FPD):

  • *Operating temperature: 350 ℃;

  • *Minor inspection measurement: P 0.2pgP/s (tributyl phosphate); S 4pgS/s (dodecanethiol)

  • Dynamic range: P 104;S 103


5. Chromatography workstation:

Contains original GCSolution Chinese or English chromatography operation software, which has the following functions:

  • Can automatically start/stopInstrument with system self diagnostic function;

  • All information (including chromatograms, quantitative calculation results, environmental settings, device settings, data processing parameters, batch tables, and report specifications) is saved in one file;

  • Assistant bar guides analysis operations;

  • User management function;

  • Effective support;

  • Event tracking support;

  • QA/QC function;

  • It can be monitored and operated through the network, and remotely accessed and diagnosed through the Internet;

  • Server based management systems can be utilized for centralized data management;


to configure:

GC-2010AF host FID detector; Split/No Split Injection Port; Single channel automatic flow controller FID auxiliary gas electronic flow controller; GC Workstation Workstation; Chinese version chromatography software; Capillary column 30m x0.32mm x0.25um non-polar capillary column; 30m x0.32mm x0.25um strongly polar capillary column; This joint must be equipped when the air compressor is paired with the pneumatic accessories; Micro injector; Detector: APC flame photometric detector, sulfur filter, phosphorus filter, auxiliary temperature control unit; ECD detector, consumables: ordinary injection pad, high-temperature injection pad, graphite gasket, split flow liner, non split flow liner, quartz cotton. Brand computer printer.

Junqi Instrument EquipmentA new batch of Agilent equipment has arrived, including Agilent 6890 gas chromatograph (equipped with single capillary FID detector; dual capillary FID detector, ECD detector); Agilent 6890N gas chromatograph single capillary FID detector; Equipped with automatic liquid sampler and sampling tower, with a hundred position sample tray; Agilent 7694 headspace sampler (44 stations fully automatic, 6 stations sample heating); Agilent 6890N+5973 gas chromatography-mass spectrometer GCMS Shimadzu QP2010 GCMS, Agilent LC-MS, AB (API) GC-MS, etc.

Before making a purchase, be sure to contact the small shopHelp you find the item you need at a suitable price

contact information 18017027751

www.junqi-sh.com

www.sh-junqi17.com

Specific configuration price, phone contact

Junqi Instrument's new chromatography equipment rental service is currently limited to customers in Jiangsu, Zhejiang, and Shanghai. The rented equipment includesAgilent 5890Pgas chromatographAgilent 6890 gas chromatograph Agilent 6890N-5973 gas chromatography-mass spectrometry Yishimitsu QP-2010 gas chromatography-mass spectrometer details consultation Junqi Instrument 18017027751

Updated on September 2, 2020 at 10:22:27

Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!