VIP member
UPC series ultra pure water machine (economical)
UPC series ultra pure water machine (economical)
Product details
UPCSeries laboratory ultrapure water machine
UPCMain performance of series ultrapure water machine instruments:
Microcomputer controlled LED color screen display, fully automatic water production
Dual outlet, capable of simultaneously producing UP water and RO water
Real time display of flushing, water production, full water, short water, and maintenance status
Automatic timed flushing, automatic water shortage protection
Automatic start-up for water intake, automatic shutdown without water use, energy-saving, easy maintenance, completely replacing traditional distilled water dispensers
UPCMain performance of series ultrapure water machine instruments:
Microcomputer controlled LED color screen display, fully automatic water production
Dual outlet, capable of simultaneously producing UP water and RO water
Real time display of flushing, water production, full water, short water, and maintenance status
Automatic timed flushing, automatic water shortage protection
Automatic start-up for water intake, automatic shutdown without water use, energy-saving, easy maintenance, completely replacing traditional distilled water dispensers
model | Ultra pure water model( UPC ) | |
scope of application | High sensitivity ICP/MC, PPT level analysis, liquid chromatography, atomic absorption, spectroscopy, mass spectrometry analysis, disease control, drug testing institutes, environmental monitoring stations, university research laboratories, and high-precision instrument water. | |
Inlet water source |
Generally, urban tap water, RO water, deionized water, distilled water, and well water have a water temperature of 5-45 ℃, TDS<200PPM, When TDS>200PPM, a water softener is required |
|
Inlet water pressure | 0.15-0.4MPa(1-5KG) | |
speed | Water production speed: 10L/h Water intake speed: 1.2L/min | |
Technology Refers to Standard |
(RO) | Conductivity<1-5 μ s/ cm@25 ℃, better than the Chinese laboratory water standard GB6682-2008 Grade II water standard |
(UP) | Conductivity ≤ 0.2 μ s/cm @ 25 ℃, resistivity 18.25M Ω .cm@25 ℃, better than the Chinese laboratory water standard GB6682-2008 Grade I water standard | |
Pyrogen content | <0.01Eu/ml | |
TOC | ≤10ppb | |
Particle content | <1/ml(≧0.22um) | |
Microbial/colony count content | <1cfu/ml | |
heavy metal | <0.1ppb | |
power supply | 220V/50Hz, 50-60W, Maximum weight 40KG | |
accessory | ◆ Water source pretreatment device (imported filter shell) ◆ Softening device ◆ Inlet pressure stabilizing device |