Tengzhou Zhongke Spectral Analysis Instrument Co., Ltd
Home>Products>Total Sulfur Gas Chromatography Analyzer for Natural Gas
Product Groups
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
    18663278921
  • Address
    No. 20, Xibeitan, Tengzhou City, Shandong Province
Contact Now
Total Sulfur Gas Chromatography Analyzer for Natural Gas
Product Introduction: "Analysis Case - Total Sulfur Gas Chromatography Analyzer in Natural Gas" is provided by Tengzhou Zhongkepu. For details, our co
Product details
brand ZKP/Zhongkepu

Total Sulfur Gas Chromatography Analyzer for Natural GasGC-2010-KT

1. Method Overview: Trace sulfur in natural gas is determined by gas chromatography, which is widely used in China,Its measurement range is:010g/m3)Using a flame photometric detector(FPD)Measurement. Adopting sample diversion technology.

2. Trace sulfur gas chromatography instrument analysis uses a highly sensitive flame photometric detector to detect organic and inorganic sulfur. Inorganic sulfur in natural gasH2SSO2Organic sulfurCOSCS2CH3STotal sulfur is the sum of all sulfur contents and can be detected0.01ppm

3. Instruments and materials: gas chromatograph (with flame photometric detector); workstation; Standard gas.

4. chromatographic column Polytetrafluoroethylene column, long1Rice.

5. Measurement steps Introduction of standard gas: Inject the standard gas twice in a row, and the relative deviation between the peak heights of the two injections should not exceed1%Otherwise, for the third time, take the values of two consecutive repetitions that pass the repeatability test, and take the average as the calibration value. Introduction of sample gas: Connect the sample gas container to the injector, using at least equivalent valves and quantitative tubes20A six way valve for blowing and cleaning gas with a sample volume of multiple times. After stopping the blowing, switch the valve and perform quantitative calculation.6.The following is the detailed configuration of the instrument:

Serial Number

product name

Specification and model

Basic configuration and parameters

quantity

1

Gas chromatograph host

GC-2010

Computer full control,Digital communication; Menu based operation; Stage programmed heating system; Automatic rear door for rapid cooling; Near room temperature operation; Four temperature controls; Fault self diagnosis, overtemperature protection, stopwatch timing, overtemperature protection and other functions are complete (including gasification chamber)

1.0tower

2

flame photometric detector

GC-2010

Sensitivity:Mt≤1×1012g(s)/sSensitivity:Dt≤Calculated by sulfur≤0.02ppm(Calculated by sulfur)

drift≤0.1mv/hrelative root mean square error≤15FPDdetection limit: Mf1×10-11g/s(Benzene)/CS2 Noise sound:0.02mvfloat move:0.1mv/h

1cover

3

Chromatography workstation

N2000

Zhejiang University adopts a dual channel systemUSBFull speed communication

1.0cover

4

chromatographic column

()

Inorganic sulfur chromatography column *Imported diatomaceous earthChromsorb(PTFE column)

1.0cover

5

chromatographic column

()

Organic sulfur chromatography column (PTFE column)

1.0cover

6

hydrogen generator

HG-1803A


Hydrogen purity:99.999%Four pole filtering, high-precision voltage stabilization. (Beijing)

1.0tower

7

Air source

Gas phase

Air purity:99.999%

1.0bottle

8

Nitrogen gas source

Gas phase

Nitrogen purity99.999%

1.0bottle

9

Standard gas

H2SSO2ppm)With316Needle Valve

1.0bottle

10

Standard gas

COSCS2CH3Sppm)With316Needle Valve

1.0bottle

11

six-way valve

GC-2010

Input pressure upper limit: 0.6MpaCentral hole of inlet and outlet air connector:3mm
Bottom plane diameter: 32mmheightIncluding a rotating handle):75mmBottom to installation panel:50mm

1.0cover

12

computer

Xpsystem

Brand computer (chromatograph)

1.0cover

13

printer

HP1020

hostx1Consumablesx1power linex1User Manualx1Random discx1warranty cardx1

1.0cover

Instrument after-sales commitment

1、 Preparation work before installation

1.The equipment we provide is of excellent quality, brand new and unused, has passed the factory quality inspection, is well packaged, and suitable for long-distance transportation.

2.Before shipment10We will notify the end user in writing of the preparation conditions for equipment installation, and upon receiving the user's written shipping request, we will ship on time.

2、 Installation and Debugging

1.We are responsible for on-site installation and debugging of the equipment provided until the system operates normally.

2.After the installation and debugging are completed, our technicians with more than three years of maintenance experience will provide training to the demand side at the installation site, so that they can independently complete various operations related to the equipment and system, as well as pay attention to precautions and handle common simple faults during use, ensuring that users can proficiently operate and maintain the instrument on a daily basis.

3、 Services provided during the warranty period

1.We guarantee to provide goods for a period of time1Annual warranty, with the warranty period calculated from the date of acceptance signature. During the warranty period, we will strictly comply with the relevant laws and regulations of the People's Republic of China regarding the provision of goodsthree guarantees"Waiting for after-sales service,Unless otherwise specified in the contract.

2.During the warranty period, due to the malfunction of our equipment, we will continue to do so24Response within hours, upon receiving notification3Within one working day, on-site service must be provided and continuous repairs must be carried out until the fault is resolved and the equipment is restored to normal.

3.We have (or entrust) * permanent maintenance institutions in China to handle all maintenance services, and have dedicated technical personnel with more than three years of corresponding work experience. Adequate supply of spare parts can meet the annual needs of users.


Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!