■ Product Usage:
Suitable for salt content in crude oil and its products, and can be used for heavy oil, residual oil, refineries and oil fields, various industrial water and discharge water
The determination of salt content can also be used for the measurement of inorganic chloride ions in the above-mentioned samples.
Applicable standards: Developed and produced according to SY/T0536 and other standards
■ Functional features:
WINDOWS operating platform, with human-machine dialogue function, easy to operate.
The screen displays parameters, measurement peaks, and measurement results, and can be stored or printed.
The electrolytic cell adopts a specially made ceramic sand core instead of the traditional anode cathode membrane, which improves the measurement sensitivity.
Test principle:
The instrument adopts the principle of microcoulometry titration. The test sample or standard sample enters the salt electrolysis cell, and the CI - in the sample reacts with the AG+in the electrolyte
The CI - and salt content of the sample can be calculated based on Faraday's law by calculating the amount of electricity consumed by the generation of AG+.
■ Technical parameters:
Measurement range: Salt content: 0.2~10000MGNACL/L Inorganic chloride ion: 0.1PPM~% content
Measurement accuracy: ≤ 1MG/L relative deviation: ± 0.2 MG/L>1MG/L relative deviation: ± 10%
Instrument usage environmental conditions: ambient temperature of 15-30 ℃, humidity ≤ 85%
Power supply voltage: AC220V, 50HZ
Dimensions: 450 × 330 × 550MM (length × width × height)