Harbin Tianlong Instrument Manufacturing Co., Ltd
Home>Products>TLXS-200 X-ray Fluorescence Sulfur Analyzer
Product Groups
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
    13351885528 13945136889
  • Address
    9th Floor, Building 12, No. 3043 Zhigu 2nd Street, Songbei District, Harbin City, Heilongjiang Province
Contact Now
TLXS-200 X-ray Fluorescence Sulfur Analyzer
TLXS-200 X-ray Fluorescence Sulfur Analyzer
Product details

Product Usage:

1. Measure the total sulfur mass percentage content in crude oil, petroleum, heavy oil, diesel, kerosene, gasoline, and other oil products:

2. Measure the total sulfur content in coal chemical products such as primary benzene:

3. Measure the total sulfur or sulfide content in solid fine powder samples, such as anode carbon blocks, petroleum coke, modified asphalt and other carbon materials:

4. Measurement of total sulfur or sulfide content in lubricating oil and petroleum additives;

5. Measurement of total sulfur or sulfide content in other liquids.

Applicable standards: GB/T 17040, GB11140, ASTM D4294 standard requirements

1. Functional Features

Wide variety of testing and wide testing range Wide variety of detection and wide detection range

General Gong

Using the fluorescence intensity ratio analysis method, temperature and pressure are automatically corrected, and the carbon hydrogen ratio (C/H0) can also be corrected:

<>

<>

<>

Users can choose the measurement result unit, pm or (i/m)%;

1 Technical parameters

1. Sulfur measurement range: 0. 0017-~5%;

2. Repeatability (r):<0.4347="" x0.="" 6446="" (x="average of two repeated test results)">0>

Reproducibility (R):<1.19182="" x0.="" 6446="" (x="two singles -">1.>

3. Sample size: Oil: 3m1~-5ml; Powder sample: 3 grams;

4. Measurement time: 30, 60, 90, 120, 150 seconds can be set arbitrarily;

5. Single sample automatic measurement measurement times: 1, 2, 3, 5, 10 can be set arbitrarily, and the average value and standard deviation will be given at the end of the measurement:

6. The instrument can store 10 calibration curves,

7. Working conditions: Temperature: 10-30 ° C: Relative humidity: ≤ 85% (30'C) ;

Power Supply: AC220V+20V,50Hz; Rated power: 50W;

8. Size and weight: length 460m X width 260m X height 260m; 11kg.

Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!