Super digital constant temperature oil bath
HH-S digital constant temperature oil bath I. Overview HH-S digital constant temperature oil bath is made of stainless steel and manufactured with tec
Product details

HH-S digital constant temperature oil bath

oneOverview

HH-S digital constant temperature oil bath is manufactured using stainless steel and technology, with digital temperature measurement. All performance indicators meet the requirements of national standards, and it has the advantages of high temperature control accuracy, strong corrosion resistance, energy saving, and long service life. It is widely used for precision constant temperature direct or auxiliary heating in scientific fields such as biology, physics, plants, chemical engineering, and environmental protection.

2、 Technical parameters

1. Power supply 220V 50Hz

2. Temperature control range: Room temperature -299 ℃

3. Constant temperature accuracy ≤± 1 ℃

4. Heating power 2000W

5. Sink size 300X300X300mm

6. Equipped with mechanical stirring function, the oil temperature rises more evenly, with a speed range of 0-3000 revolutions per minute

3、 Usage method

Operating instructions

1. To set the temperature, press the SET button. Pressing the SET button will flash the data on the digital display in the lower row (normal temperature measurement in the upper row), indicating that the instrument has entered the temperature setting state. Press the Δ button to increase the set value, press the ▽ button to decrease the set value, and continue pressing the increase or decrease button will quickly change the set value. Press the SET button again to return the instrument to normal working state. The temperature setting is complete.

2. The correction of sensor errors can be made when confirming that the displayed value on the instrument is not the correct measurement value.

Press the SET key for 3 seconds to enter the inner menu of the instrument panel. After selecting the parameter SC, use the Δ or ▽ keys to set this parameter. The range for correcting sensor errors is+50 to -50. After the correction is complete, press the SET button for 3 seconds to exit. The correction value of the instrument is 0 when it leaves the factory. When using it, it is necessary to prevent the instrument that displays correctly from being corrected incorrectly.

4、 Setting and Explanation of User Working Parameters

Press the SET button for 3 seconds to enter the inner layer parameter setting state of the instrument.

Parameter code

Parameter meaning

describe

set range

AL

Upper limit alarm

Set upper limit alarm value

full scale

Hy

Main output hysteresis

Main output action insensitive area

0-50

P

Proportional band

If it is 0, it is positional control

1℃-200℃

I

integral time constant

If it is 0, cancel the integral control function

0-999 seconds

d

derivative time constant

If set to 0, cancel the differential action

0-999 seconds

t

Proportional cycle

Adjust the output action cycle

1-200 seconds

Sc

Measurement value bias

Sensor error correction

-50℃-50℃

Lck

Parameter lock

0 is not locked; 1. Lock parameters other than the set value; Lock all parameters

0-2

Parameter significance

AL overtemperature alarm parameter, when the temperature exceeds the AL value, the ALM indicator light will turn on.

HY main output hysteresis prevents frequent relay actions.

P is the proportional band, which refers to the proportional control value. The instrument is centered around the set value and the P value is distributed on both sides of the set point. Within the P value range, the instrument operates in a P I D mode.

I is used to eliminate static error, too small can cause temperature overshoot. Usually 80-180 seconds.

D overshoot control, too large can cause fluctuations, too small can slow down the system. Usually 10-40 seconds.

T is the heating output cycle period, which is the working cycle period of the relay. A small T value has good control effect, but a too small T value will reduce the lifespan of the relay due to frequent operation. Generally, the T value is taken as 20-60 seconds.

SC error correction.

LCK parameter locking to prevent misoperation.

5、 Precautions

1. Before use, please remember to add oil at a height of 1cm above the bottom lining plate before turning on the power. Do not turn on the power before adding oil to the box to avoid burning out the electric heating tube.

2. Remember to cut off the power before draining the oil.

3. Properly grounded for safety.

Our products are covered by a one-year warranty from the date of delivery, with lifetime maintenance (additional fees may apply for damages caused by human error or improper use). Please feel free to use them.