SW-CJ-1F series single person double-sided, vertical air supply Ultra clean workbench Purification workbench Clean workbench Ultra clean table Purification table Clean table bioclean bench
SW-CJ-1F series single person double-sided, vertical air supply Ultra clean workbench Purification workbench Clean workbench Ultra clean table Purific
Product details

SW-CJ-1E/ SW-CJ-1FPurification workbench is a local purification unit used in the fields of medicine and health, biotechnology, scientific experiments, etc. for sterile and dust-free clean environments.

technical parameter

model

SW-CJ-1ESingle person double-sided purification table (all stainless steel shell)Applicable number of people: Single person single-sided

SW-CJ-1FSingle person double-sided vertical air supply ultra clean/purification workbenchApplicable number of people: Double single-sided

Cleanliness level

100Grade @ ≥ 0.5um (Federal 209E)

100Grade @ ≥ 0.5 μ m Federal 209E

cfu

≤ 0.5 pieces per dish Time (Φ 90mm culture dish)

≤ 0.5 per dish per hour (in a 90mm culture dish)

Average wind speed

00.25m-0.45/m/s (fast and slow dual speed)

0.25-0.45m/sFast slow dual speed)

Vibration half peak value

≤ 5 μ m (X.Y.Z direction)

≤ 5 μ m (X.Y.Z direction)

noise

≤62dB(A)

≤62dBA

illumination

≥300Lx

≥300Lx

Power Supply

ACSingle phase 220V/50Hz

AC,Single phase 220V/50HZ

Maximum power consumption

130W

0.4KW; 0.8KW

weight

<85kg

150KG

Specification and quantity of high-efficiency filters

695×505×38×①

865×558×50×①?

Specification and quantity of fluorescent/ultraviolet lamps

8W×① 10W×①

20Wx①/20Wx①

Applicable number of people

Single person single-sided

Double single-sided

External dimensions

850×580×1600mm

1050×680×1600

Size of workspace

700×500×500mm

870×650×520


maintain

1According to the actual situation, regularly remove and clean the primary filter, with a cleaning cycle of generally 3-6 months. If not cleaned for a long time, dust accumulation will lead to insufficient air intake and reduce the cleaning effect.

2When the ideal cross-sectional wind speed cannot be achieved after normal replacement or cleaning of the primary air filter, the operating voltage of the fan should be adjusted to achieve the ideal uniform wind speed.

3Generally, when the operating voltage of the fan is adjusted to the highest point after 18 months of use and still cannot reach the ideal wind speed, it indicates that the high-efficiency air filter has accumulated too much dust (the filter holes on the filter material are basically blocked and need to be updated in a timely manner). The service life of a high-efficiency air filter is generally 18 months.

4When replacing the high-efficiency air filter, attention should be paid to the correct model, specifications, and dimensions (as configured by the original manufacturer), install the device according to the direction of the arrow, and ensure that the surrounding seal of the filter is sealed and there is no leakage.

General fault causes and troubleshooting methods

Fault phenomenon

reason

Exclusion method

The main power switch cannot be connected and automatically trips

1The fan is stuck, causing the motor to stall, or there is a short circuit in the circuit.

1Adjust the position of the fan shaft, or replace the impeller and bearings, and check if the circuit is intact.

2Check the insulation resistance of the circuit and components to the casing point by point according to the circuit diagram, and repair the insulation fault area.

Low wind speed

1. Excessive dust accumulation in the primary filter

2. High efficiency filter failure

1. Clean the primary filter.

2. Replace the high-efficiency filter.

The fan is not turning

1. The contactor is not working

2. The fuse core of the fan power supply has blown

1. Check if the contactor circuit is correct.

2. Replace the melt core.

The fluorescent light is not on

1. Lamp or relay damaged

2. The fuse of the lamp power supply has blown

1. Replace the light tube or relay.

2. Replace the melt core.