Portable online gas analyzerTechnical Parameter
Temperature control indicators
Column box: 5 ℃ to 400 ℃ at room temperature with an accuracy of ± 0.1 ℃
Injector and detector: above room temperature 6 ℃~399 ℃ accuracy ± 0.1 ℃ Program heating stage: fifth order
Program heating rate: 0.1 ℃~40 ℃/min at room temperature of 6 ℃~200 ℃
0.1 ℃ to 20 ℃/min when the temperature exceeds 200 ℃
Detector indicators
Thermal conductivity detector (TCD):
Sensitivity S ≥ 4000mv. ml/mg (n-hexadecane)
Noise ≤ 20UV
Linear range ≥ 104
Nitrogen phosphorus detector (NPD):
Sensitivity: Nitrogen Mt ≤ 5 × 10-11g/s
Phosphorus Mt ≤ 2 × 10-12g/s
Noise N ≤ 2 × 10-13A
Electron Capture Detector (ECD):
Sensitivity Mt ≤ 1 × 10-13g/ml (r-666)
Noise ≤ 15UV
Linear range ≥ 5 × 103
Radiation source Ni63
The maximum operating temperature is 350 ℃
Flame photometric detector (FPD):
Method: S or P filter
Sensitivity: 5 × 10-11g (s)/s (S in thiophene)
1.4 × 10-12g (p)/s (P in methyl parathion)
Portable online gas analyzerexecuteStandard:
GB 10410.3-89 "Gas Chromatography Analysis of Liquefied Petroleum Gas Components"
SH/T 0230-92 "Determination of Composition of Liquefied Petroleum Gas (Chromatography Method)"
Application scope:
Industrial gas application
Application of Civil Gas
Application of metallurgical reducing gas
Chemical synthesis raw material gas
Applications in other fields such as coal to hydrogen production