Online methane non methane total hydrocarbon analyzer, gas chromatograph manufacturer
The manufacturer of the online methane non methane total hydrocarbon analyzer gas chromatograph adopts a hydrogen flame ionization detector, which has
Product details
Online methane non methane total hydrocarbon analyzer, gas chromatograph manufacturer
The hydrogen flame ionization detector has high sensitivity and good stability, mainly used for online analysis of methane and non methane total hydrocarbons.
Characteristics of methane non methane total hydrocarbon analyzer:
1. Can be connected to external temperature, humidity, and pressure sensors.
2. The temperature, humidity, and pressure data of the tested environment can be manually entered.
3. Support temperature compensation to obtain accurate data in cold environments.
4. Support humidity compensation to obtain accurate data in high humidity environments.
5. Customizable preheating time for sensors.
6. Multiple calibration points can be established to make the detection results more accurate.
7. It can roll back the calibration history to prevent inaccurate detection caused by misoperation.
8. Temporary zero point technology can detect low concentration trace gases and is very suitable for use as a gas leak detector.
Online methane non methane total hydrocarbon analyzer, gas chromatograph manufacturerInstructions for use:
1、 Daily monitoring of analytical instruments, gas processing systems, high-temperature probes, cooling water, and purge gas
Conduct inspections on complete sets of system equipment, including:
1. Check if the instrument displays any fault alarms
2. Observe the changes in sample gas flow rate
3. Condenser operating status
4. Peristaltic pump drainage or water accumulation in the reservoir
5. Whether the operating status of the analytical instrument and the indicated values of the analytical instrument are normal
6. Heating condition of sampling probe
7. Compressed air pressure for blowing (greater than 0.3MPa) and manual drainage of the front stage air-water separator
8. Does the humidity alarm membrane filter sound an alarm, and does the high-precision filter have water or dust ingress
9. Usage of standard gas
10. Analyze the temperature inside the instrument room
11. Other abnormal situations during operation, etc
2、 Regularly (within one month) perform zero and range calibration on the measured components of analytical instruments, and keep records. When calibrating analytical instruments, the calibration of standard gas cylinders is crucial