New intelligent touch upgrade product
Technical features:
1:Support testing of multiple parameters such as oxidation induction period, melting point, cold crystallization, solidification, glass transition temperature, specific heat capacity, etc., with one click operation by customers and fully automatic software operation.
2:A brand new sensor sample holder with a fully enclosed structural design to prevent items from falling into the furnace body.
3:The instrument complies with the following national standards, but is not limited to them:
GB/T 19466.2-2004/ISO 11357-2: 1999 Part 2:Determination of glass transition temperature;
GB/T 19466.3-2004/ISO 11357-3: 1999 Part 3:Measurement of melting and crystallization temperature and enthalpy;
GB/T 19466.4-2016/ISO 11357-4: 1999 Part 4:Determination of specific heat capacity;
GB/T 19466.6- 2009/ISO 11357-3: 1999Part 6:Determination of oxidation induction time (isothermal OIT) and oxidation induction temperature (dynamic OIT) during the oxidation induction period.
4:Embedded with a 10 inch ultra large industrial grade computer, no need for a separate computer, integrated device,Rich display information, including set temperature, sample temperature, oxygen flow rate, nitrogen flow rate, differential thermal signal, various switch states, and zero flow rate.
5:The instrument comes standard with four USB ports on the right side, making it convenient for users to connect external devices such as mice, keyboards, printers, and USB drives, making it easy for users to use.
6:The internal USB communication interface of the instrument is more stable and versatile compared to the original instrumentReliable and uninterrupted communication, supporting self recovery connection function.
7:The digital gas quality flowmeter automatically switches between two atmospheric flow rates, with fast switching speed and short stabilization time.
8:Standard samples are provided as standard, making it convenient for customers to calibrate the constant temperature coefficient.
9:Pre installed software, no computer compatibility issues,Reduce connection issues caused by computer replacement or driver installation errors.
10:Support user self programming programs to achieve full automation of measurement stepsThe software provides dozens of instructions, and users can flexibly combine and save each instruction according to their own measurement steps. Simplify complex operations into one click operations.
Product application scope:
1:Differential scanning calorimeter measures the relationship between temperature and heat flux related to the internal thermal transformation of materials, and has a wide range of applications, especially in material research and development, performance testing, and quality control. The characteristics of materials, such as glass transition temperature, cold crystallization, phase transition, melting, crystallization, product stability, curing/crosslinking, oxidation induction period, etc., are all research areas of differential scanning calorimetry.
2:Application scope of differential scanning calorimeter: determination of curing reaction temperature and thermal effect of polymer materials, phase transition temperature and thermal effect of substances, determination of crystallization and melting temperature and thermal effect of polymer materials, and glass transition temperature of polymer materials.
Technical parameters:
1:Temperature range: Room temperature~500 ℃
2:Temperature resolution: 0.01 ℃
3:Heating rate: 0.1-80 ℃/min
4:Temperature control method: heating up, constant temperature (fully automatic program control)
5: DSCRange: 0 to ± 500mW
6: DSCResolution: 0.001mW
7: DSCSensitivity: 0.001mW
8:Working power supply: AC 220V 50Hz or customized
9Atmosphere control gas: nitrogen oxygen (instrument automatically switches)
10Gas flow rate: 0-200mL/min
11:Gas pressure: 0.2MPa
12:Gas flow accuracy: 0.2mL/min
13:Optional crucible: Aluminum crucible, diameter 6.7 * 3mm
14:Parameter standard: equipped with standard substances (indium, tin, zinc, sapphire), users can calibrate temperature and enthalpy by themselves
15:Data interface: Standard USB interface (built-in interface, no external connection required)
16:Display mode: 10 inch industrial computer touch display, can be connected to external mouse, keyboard, USB flash drive, printer