1brief introduction

QNix4200andQNix4500These two models are designed in an integrated manner, requiring only zero adjustment and no calibration, making them extremely easy to use. of whichQNix4200As a magnetic thickness gauge, it can be used to measure non-magnetic coatings and coatings on magnetic substrates such as steel and iron;QNix4500For both magnetic and eddy current dual-use thickness gauges, it can not only be used to measure non-magnetic coatings on magnetic substrates such as steel, but also to measure non-conductive coatings on non-magnetic metal surfaces such as aluminum, copper, stainless steel, such as paint layers, oxide films, phosphating films, etc. These two models are easy to operate, portable, and have high precision, which are loved by a large number of users.

尼克斯4500

2、 Technical parameters

Magnetic matrix FeProbe(4200/4500

Non-magnetic matrix NFeProbe(4500

measuring range Fe:0-3000μm/NFe:0-3000μm

accuracy 0-50μm:+/-1μm

50-1000μm:+/-1.5%reading

1000-3000μm+/-3%

Minimum contact surface 10×10mm

Minimum curvature radius Convex surface:3mm

concave:25mm

Minimum body thickness Fe:0.2mm/NFe:0.05μm

Temperature compensation range -10-60

measuring range 0-50

display LCDliquid crystal

probe Ruby fixed type

power supply 2×1.5Vdry cell

size 100×60×27mm

weight 105g

尼克斯4500

3、 Precautions and common questions
1.Measurement should be point contact, and it is strictly prohibited to place the probe on the surface of the object being measured and slide it.

2.appearINFIWhen measuring, it may be due to incorrect selection of the measuring substrate. Please select the measuring substrate according to the fourth itemFENFEOr automatic recognitionFE/NFEIf the issue cannot be resolved, please perform the third zeroing process.

3.Due to different thickness units used in different countries and regions, the products sold in ChinaQnix4500/4200The type used isμmTherefore, British units were blockedmilSometimes activated due to improper operationmilUnit, at this time, the user only needs to reset. The reset process is to first remove the battery, then press and hold the red button to install the battery, and release the red button after the reset is complete.

尼克斯4500

4、 After sales service

Imported instruments enjoy original manufacturer benefitsOne year warranty, lifetime maintenanceService

You can enjoy our company's trade in program.

Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!