1. Product configuration
Please confirm if the packaging box of the PH-1186 detector you purchased is complete. If there is any packaging damage or shortage of any accessories, please contact the dealer as soon as possible. The configuration is as follows.
Standard configuration
◇ One SHYYPH-1186 instrument
◇ One PH or ORP electrode (note: included in one set)
◇ Tighten two locking strips
◇ One user manual
◇ Two packs of pH calibration buffer powder (6.86, 4.00)
Optional Accessories
◇ Electrode installation bracket
◇ Electrode flow cup
2. Product Introduction
SHYYPH-1186 is a new generation intelligent water quality detector independently developed by our company, with patent rights. It has the characteristics of Chinese menu operation, high stability, superior repeatability, and multifunctionality, and can accurately measure the pH value, oxidation-reduction potential (ORP), and temperature value in the solution. Widely used for online detection of pH values in solutions such as environmental protection, sewage treatment, thermal power, aquaculture, food processing, printing, metallurgy, pharmaceuticals, fermentation, chemical engineering, and tap water.
Main features:
◇ Large screen dot matrix LCD display, Chinese menu operation, and simultaneous display of multiple parameters.
◇ Automatic temperature compensation function.
Communication function: Equipped with RS-485 communication interface (partially compatible with MODBUS protocol) (optional), capable of converting to RS-232. The pH value corresponding to the 4-20 mA current output can be set arbitrarily.
◇ Optional upper computer online acquisition software from our company for data collection and processing.
◇ Notepad: PH/ORP mode can store 50 measurement data respectively, and users can record measurement data as needed.
◇ Historical curve (optional): PH/ORP mode can record 30 days of historical data (1/10 min) respectively, and has a query function that can reset the starting recording point.
◇ Automatically recognize standard solution and restore factory settings.
Watchdog function: Ensure that the instrument panel does not crash.
◇ Power failure protection>10 years.
The core components come from well-known foreign brands.
3. Technical indicators
1. Measurement range: -2.00~16.00PH, -1999~+1999mV (ORP), -10~130.0 ℃
2. Resolution: 0.01PH, 1mV, 0.1 ℃
3. Accuracy: ± 0.02 PH, ± 1mV, ± 0.3 ℃
4. Stability: ≤ 0.01pH/24h
5. PH standard solution: 6.86/4.00/9.18??? 7.00/4.00/10.01
6. Temperature compensation: 0-99.9 ℃ (pH)
7. PH correction range: zero point ± 1.45PH; slope ± 30%
8. Control interface: Two sets of ON/OFF relay contacts, divided into high point and low point alarm control.
9. Signal isolation output: 4-20mA isolation protection output
10. Relay: Relay hysteresis can be set arbitrarily, relay load 10A 220VAC
11. Working conditions: ambient temperature of 0-60 ℃, relative humidity ≤ 90%
12. Input impedance: ≥ 1 × 1012 Ω
13. Output load: load<500 Ω (0-10mA), load<750 Ω (4-20mA)
14. Working voltage: 220VAC 10%, 50/60Hz
15. Size: 96 × 96 × 115? mm
16. Hole size: 91 × 91mm
17. Weight: 0.9Kg.