The HORIBA oil analyzer OCMA500 button enables oil extraction, measurement, and drainage, making the operation more convenient than before. After samp
Product details

Product Introduction

HORIBA Oil Analyzer OCMA500characteristic

1. One click simple operation
After sample injection, press the START button to automatically perform actions such as oil extraction, measurement, drainage, and opening/closing of the drainage valve. Simple operation enables fast measurement. Moreover, the design of color LCD display screen and extraction tank with backlight provides better operability for the analyzer.
2. Environmentally friendly and lower measurement costs
OCMA500 is an infrared oil analyzer that extracts oil from samples using extractant S-316 and analyzes oil concentration using the principle of infrared absorption. The amount of extractant used has decreased compared to before, achieving a reduction in environmental load and effectively reducing measurement costs.
3. Diversified uses
1) Drainage
Measurement of oil content in various types of factory drainage (steel, industrial waste, petrochemicals, food)
Measurement of oil content in wastewater from sewage treatment plants
Monitoring of oil content in the bottom sediment and curved drainage of cruise ship cabins used for ocean transportation
Measurement of oil content in wastewater from waste oil treatment equipment and performance evaluation of memory oil separation equipment
2) Environment
Environmental water quality investigation based on environmental standards
Water quality measurement around motor vehicle inspection sites and gas stations
Crude oil, measurement of oil content in the cleaning and drainage of storage tanks in oil bases
Investigation of oil dispersion during accidents on cruise ships, oil platforms, and petrochemical factories
3) Parts
Measurement of residual oil content in parts cleaning engineering
Evaluation of oil content in parts cleaning solution
Evaluation of Oil and Fat Cleaning for Semiconductor Coating Contact Surface

4. Color LCD display
Adopting a 3.5-inch color LCD display, the operation is more convenient and data reading is more convenient
5. USB interface
Compared to using USB memory for data storage, using a computer for data management will be easier
6. Backlight design for extraction tank (technology)
The extraction tank is equipped with LED lights, which can provide illumination and facilitate the confirmation of the stirring and separation status of the extraction agent and sample
7. Unit optional functions
Input measurement conditions and select appropriate units (mg/l, mg/kg, mg/g, mg/pc) based on the measurement purpose
8. Due to the absence of an extraction agent evaporation process, not only is the operation simple, but low boiling point oil can also be measured
Due to the n-hexane extraction method, the extractant needs to be evaporated, which can cause low boiling point oil components to also evaporate. OCMA-500 does not require an evaporation process as it can effectively prevent the volatilization of low boiling point oils such as toluene and gasoline.


Optional: Solvent Regenerator SR-305
SR-305 is a solvent regenerator that can exhibit good operability and regeneration ability through column layers of activated carbon and activated alumina.

HORIBA Oil Analyzer OCMA500Technical Parameter
Application: Measurement of oil content in drainage, environment, parts, and soil analysis
Measurement principle: Extraction with extractant, non dispersive infrared analysis
Detector: Thermoelectric sensor
Measurement range: 0-200mg/L
Resolution: 0-99.9mg/L: 0.1mg/L
100-200mg/L:1mg/L
Repeatability: 0-9.9mg/L: ± 0.2mg/L ± 1 digit
10.0-99.9mg/L: ± 2.0mg/L ± 1 position
100-200mg/L: ± 4mg/L ± 1 digit
Display screen: 3.5-inch backlit color LCD display screen
Measurement: Fully automatic measurement, also available with manual option
Calibration: Zero point, range calibration: Automatic or manual calibration after injecting liquid
Extraction agent: S-316
Sampling volume: 2:1 (sample: extractant)
Extraction method: Built in extraction tank
Function: Can store 300 sets of data, self diagnosis, measurement stability display, clock, mixing tank with backlight, unit conversion
Output: USB storage output
Environmental temperature: 0-40 ℃
Power supply: 100-240V AC ± 10%, 50/60Hz
Rated power: Approximately 60VA at AC 100V, approximately 90VA at AC 240V
Dimensions: 200 (W) * 313 (D) * 342 (H) mm
Quality: 7Kg
External output: USB storage output
Measurement process: After injecting liquid, automatic measurement (automatic program switching), and manual measurement

Standard accessory
Filter element: Water separation filter element diameter 40mm, 5 pieces
Pipette: 2.5ml polyethylene material
Power cord: 1 220V power cord
B Heavy Oil: 10ml

Microinjector: 25 μ L

Instruction manual: OCMA500 user manual