GMS980 high-precision Gaussian meter calibration system |
(GMS980High precision Tesla meter calibration device)
|
|

|
sketch:
|
The GMS980 high-precision Gaussian meter calibration system is a high-precision, powerful, and intelligent Gaussian meter calibration system. This system adopts our company's high-precision fully digital Gaussian meter, high-performance, high-precision, high-power DC power supply, and high uniformity electromagnet to form a high-precision, high stability magnetic field generation platform with closed-loop control. Through software and hardware platforms, intelligent real-time display of preset values and magnetic field fluctuation graphics and values. The precision of the magnetic field generation platform is as high as ± 0.001% of the reading; Magnetic field generation degree: up to 2.4T; Magnetic field stability: 10ppm/8 hours. This system fully complies with the magnetic field calibration standard "JJG242-1995 Tesla Meter Calibration Regulations" of the Chinese Academy of Metrology, and is suitable for calibration by various metrology departments, as well as various magnetic research in scientific research institutions and military enterprises.
The GMS980 high-precision Gaussian meter calibration system has an accuracy of up to 5ppm and can calibrate Gaussian meters with an accuracy exceeding 0.01%.
|
|
Soft and Hard Platforms
|
|
1. High precision Gaussian meter, highest accuracy5ppm;
2. High performance, high-precision, high-power DC power supply, stability:10ppm / 8Hour;
3. High uniformity electromagnet:Uniformity can reach10ppm /cm³;
4. Various supporting software and high-precision adjustment brackets and other accessories.
|
|
|
|
Detailed description of system composition
|
1. High-precision Gaussian meter |
There are three types of high-precision Gaussian meters:G401The accuracy is0.04%)TheG501The accuracy is0.01%)And nuclear magnetic resonance Gaussian meterPT2025(accuracy5ppm)、PT2026(accuracy5ppm)It is mainly used in measurement departments, scientific research institutions, military industry, shipyards and other large manufacturers. |
Gauss meter |
G401 |
G501 |
range |
50KG(5T) |
30KG(3T) |
Highest resolution |
0.1mG(10nT) |
1mG(0.1μT) |
Basic DC accuracy |
±0.04% |
±0.01% |
Frequency range |
DC- 30kHz |
DC- 20kHz |
temperature coefficient |
<±50ppm>±50ppm>℃ |
<±50ppm>±50ppm>℃ |
communication interface |
USB |
USB |
Display digits |
complete6Position display |
complete6Position display |
Display |
7Inch color resistive touch screen |
7Inch color resistive touch screen |
Display units |
Gaussian(G)Tesla(T)Ampere per meter(A/m) |
Gaussian(G)Tesla(T)Ampere per meter(A/m) |
|
|
NMR Gauss meter |
PT2025 |
PT2026 |
range |
0.043T to 13.7T |
0.038T to 22.8T |
resolution |
0.1μTperhaps1Hz |
±0.1Hz |
accuracy |
±5ppm |
±5ppmNot affected by temperature |
Frequency range |
1.9MHz- 90MHz |
1MHz- 1GHz |
communication interface |
RS-232-CandIEEE-488 GPIB |
USB/USBTMC, Ethernet / VXI-11
IEE488.2; SCPI
|
power supply |
110/220V 50/60Hz, 40VA |
90-255VAC 50/60Hz, 50W |
Display units |
TorMHz |
TorMHz |
|
|
2. Model 9100High performance, high-precision, high-power DC power supply |
|

|
|
9100The high-performance, high-precision, and high-power DC power supply is a device that utilizes current/High precision and high-power DC power supply controlled by voltage, with current stability of up to10ppm/8Hours, high cost-effectiveness, excellent performance. Power up to12KW, with high efficiency93%, lightweight only32KGSuitable for power supply scenarios with high current, small size, and low cost.
Compared with traditional hard switch technology,9100The high-precision and high-power DC power supply has excellent designs such as high-efficiency bridge rectification, phase modulation, and zero voltage switching power supply. The whole machine is designed and manufactured in Europe.
|
|
|
Electromagnetic power supply |
9100 |
drift(8Within hours) |
±10ppm |
Line adjustment |
±10% slow, T > 1 min |
±5ppm |
±1% fast, T > 3 m sec. |
±5ppm |
Load regulation(± 10%resistance variation) |
±5ppm |
Output ripple and noise(voltage–peak-to-peak) |
< 100mv='' @='' 0-100=''> |
temperature coefficient |
ambient temperature15 – 40° C |
5ppm/°C |
ambient temperature15 – 30° C |
1ppm/°C |
Current setting resolution |
20 bit DAC |
Current stability accuracy |
± 10ppm |
Absolute calibration current |
-0 / +400ppm |
Current readback resolution |
16 bit ADC |
Voltage readback resolution |
16 bit ADC |
Current control range(Range setting) |
1 - 100% |
Uniform speed(0 – 100%) |
0.1 - 10 sec (adjustable) |
Current loop bandwidth |
2 - 100 Hz |
Voltage loop bandwidth |
>200 Hz |
Isolation test voltage (output as an example) |
1 kV |
Warm-up time(Cooling and standby) |
30minute;15minute |
ACpower input |
360~440VAC (400V ± 10%),Three phase four wire system,47~63Hz |
ACPower input efficiency |
AC input voltage90%~93%unequal |
DCPower Output |
voltage60V, optional current50A - 100 A - 150 A - 200 A |
|
|
model selection |
Unipolar, 380VACThree phase four wire input,water-cooling |
9100-50A-U |
50A/60VUnipolar, including: local control panel,RS232Communication interface, water-cooled |
9100-100A-U |
100A/60VUnipolar, including: local control panel,RS232Communication interface, water-cooled |
9100-150A-U |
150A/60VUnipolar, including: local control panel,RS232Communication interface, water-cooled |
9100-200A-U |
200A/60VUnipolar, including: local control panel,RS232Communication interface, water-cooled |
Bipolar, 380VACThree phase four wire input,water-cooling |
9100-50A-B |
±50A/±60VBipolar, including: local control panel,RS232Communication interface, water-cooled |
9100-100A-B |
±100A/±60VBipolar, including: local control panel,RS232Communication interface, water-cooled |
9100-150A-B |
±150A/±60VBipolar, including: local control panel,RS232Communication interface, water-cooled |
|
|
3. High uniformity electromagnet |
SBThe Type Electromagnetic Magnet is a new generation of double yoke and double adjustable structure electromagnetic magnet from Koehler Technology GmbH in Germany.SBType electromagnet, its magnetic field air gap is adjustable in both directions, the double yoke structure forms a closed magnetic circuit, has good rigidity, and the magnetic field direction is horizontal; Yoke inclined type45The angled seat is convenient for taking and placing samples, as well as for combining with other equipment.
SBThe design of the electromagnetic coil with a reasonable magnetic circuit and pole size100mm—250mm, weight from200KG—1920KGTo form a series of products. The electromagnetic coil is finely processed to ensure the coaxial alignment of the pole and the maximum number of ampere turns of the wire package. It can be condensed to the smallest volume, and can obtain the strongest magnetic field per unit volume of the wire package for every kilowatt of excitation power. It can also ensure a large smoothness around the working air gap. The electromagnet can be naturally cooled or internally cooled with water cooling to ensure that it can be used at normal temperature.SBThe electromagnetic coil has a wide range of air gap adjustment, easy replacement of pole heads, and is suitable for various magnetic measurements.
|

|
|
electromagnet |
SB-100 |
SB-130 |
SB-175 |
SB-200 |
SB-250 |
Maximum magnetic pole diameter(mm) |
100 |
130 |
175 |
200 |
250 |
Adjustable range of air gap(mm) |
0-120 |
Magnetic field(T) |
>2.0T |
>2.0T |
>2.0T |
>2.0T |
>2.0T |
power(KW) |
2.0 |
3.8 |
6.5 |
10.0 |
12.0 |
weight(KG) |
200 |
430 |
630 |
930 |
1920 |
|
magnetic field homogeneityDetermined by the following factors: electromagnetic power, magnetic pole diameter, and air gap. By selecting suitable parameters, the magnetic field uniformity can reach one hundred thousandth/Cubic centimeters.
|
|
4Various supporting software and high-precision adjustment brackets and other accessories
|
4.1 Supporting software (upper computer software)
The software can set the DC power supply current value and generate a controlled electromagnetic uniform zone magnetic field by controlling the DC power supply current output; The software can set the target magnetic field value, achieve the target value after the system works, and automatically input it into the test table. The measured data needs to be manually input. After the tester completes the test from low field to high field, the software automatically generates a measurement table and measurement results. The testing standards are formulated according to the standards of the Chinese Academy of Metrology. Test data and tables can be stored in the database for archiving. Support printing test results.
4.2 Adjustment bracket (displacement platform)
The displacement platform is made of non-magnetic materials. Users can manually do itX,Y,ZMovement in direction,Move the probe steadily to a certain position and fix it. The positioning accuracy is0.1Millimeters. Through the displacement platform, the probe can be accurately positioned in the uniform magnetic field area, eliminating the influence of position and angle.
4.3 Water circulation cooling box: matched with the selected DC power supply.
|
|
Attachment:High precision magnetic field calibration system
|
|
System type |
Magnetic field calibration range |
Main parameters and configuration |
DC magnetic field calibration system |
high range |
◆Range:10G - 20KG
◆Equipment composition:175mmElectromagnetic coil, nuclear magnetic resonance(NMR)Gaussian meter,10ppmHigh precision Degree power supply degree power supply
◆Accuracy:50ppm
|
Medium range |
◆Range:1G-1000G
◆Equipment composition: high-precision solenoid,6 A half multimeter,10ppmHigh precision power supply,Standard Gaussian meter
◆Accuracy:0.01%
|
low range |
◆Range:1μG – 1G
◆Equipment composition: large shielding bucket, three-dimensional Holmz coil,1μAHigh resolution and high-precision electrical Source, standard flux gate Gaussian meter
◆Accuracy:0.1%
|
Communication magnetic field calibration system |
high range |
◆Range:1G – 2000G
◆Frequency:10Hz-3KHz
◆Equipment composition:150ACommunication coil,6 Half multimeter, high-precision and high-power3KHzAC power supply, standard AC Gauss meter
◆Accuracy:0.1%
|
Medium range |
◆Range:0.1G – 150G
◆Frequency:10Hz-30KHz
◆Equipment composition:20ACommunication coil,6 Half multimeter with high precision30kHzalternating current Source, standard communication Gaussian meter
◆Accuracy:0.1%
|
low range |
◆Range:1mG - 1G
◆Frequency:10Hz-30KHz
◆Equipment composition: Holmz coil,6 Half multimeter with high precision30kHzalternating current Source, standard AC flux gate Gaussian meter
◆Accuracy:0.1%
|
|
|
Attached are other magnetic measurement product parameters |
Provide professional magnetic field measurement solutions |
Strong magnetic field measuring instrument(0 - 30T) |
model |
principle |
type |
number of axles |
DC accuracy |
maximum range |
Highest resolution |
G91 |
hall |
handheld |
1 |
2% |
3T |
10μT |
G92 |
hall |
handheld |
1 |
1% |
10T |
10μT |
G93 (Equipped with a strong magnetic field probe) |
hall |
handheld |
3 |
1% |
30T |
1μT |
G201 |
hall |
Desktop computer |
1 |
0.2% |
5T |
10nT |
G401 |
hall |
Desktop computer |
1 |
0.04% |
5T |
10nT |
G403 |
hall |
Desktop computer |
3 |
0.1% |
3T |
1μT |
GSP301A |
hall |
Analog transmitter |
1 |
1%, 0.2%, 0.04% |
10T |
Output:±10V |
GSP301D |
hall |
Digital transmitter |
1 |
1%, 0.2%, 0.04% |
10T |
1μT |
GSP303A |
hall |
Analog transmitter |
3 |
1%, 0.5%, 0.25%, 0.1% |
2T |
Output:±10V |
GSP303D |
hall |
Digital transmitter |
3 |
1%, 0.5%,0.1% |
2T |
1μT |
Weak magnetic field measuring instrument(0- 1mT) |
Weak magnetic field Gaussian meter |
model |
principle |
type |
number of axles |
DC accuracy |
maximum range |
resolution |
GMR50 |
reluctance |
handheld |
1 |
1% |
1mT |
10nT |
G93 (Equipped with low field probe) |
reluctance |
handheld |
3 |
1% |
1mT |
10nT |
GF603 |
fluxgate |
handheld |
3 |
0.5% |
1mT |
0.1nT |
GF633 |
fluxgate |
Desktop computer |
3 |
0.1% |
100μT |
0.01nT |
GFP703 |
fluxgate |
Smart Transmitters |
3 |
0.5% |
1mT |
0.1nT |
fluxgate sensor |
model |
principle |
type |
number of axles |
Linearity |
output |
maximum range |
noise level |
F901 |
fluxgate |
Economic type |
3 |
<> |
±10V,single-ended |
1000μT |
10 to 20pT |
F902 |
fluxgate |
low noise |
3 |
<> |
±10V,single-ended |
100μT |
<> |
F903 |
fluxgate |
low-power consumption |
3 |
<> |
±3V,finite difference |
200μT |
10 to 20pT |
F904 |
fluxgate |
low-power consumption&noise |
3 |
<> |
±3V,finite difference |
100μT |
<> |
F53 |
fluxgate |
High temperature type |
3 |
<> |
±10V,finite difference |
100μT |
≤300pT@175℃ |
Communication magnetic field measurement(30Hz- 1MHz) |
model |
principle |
type |
number of axles |
AC accuracy |
maximum range |
frequency response |
GA500 |
alternating current coil |
handheld |
1 |
1% |
3mT |
30Hz- 1MHz |
GA1000 |
reluctance |
handheld |
3 |
2% |
400μT/ Axis |
30Hz- 1MHz |
Magnetic field testing system |
GAS3000 |
Array magnetic field detection system(Multi point synchronous testing system for magnetic field)
According to the configuration, the system can monitor simultaneously2~128The magnetic field of the road
|
Magnetic field calibration system |
DC magnetic field (High precision50ppm) |
ac magnetic field (High precision0.1%) |
·Low range:0.1nT- 0.1mT ·Medium range:0.1- 100mT ·High range:1mT- 2T |
·Low range:0.1μT- 0.1mT ·Medium range:0.01- 15mT ·High range:0.1- 200mT |
remarks:1T= 10KG; 1mT= 10G; 1μT= 10mG; 1nT= 10μG |
|
|