Guiyang Xuetong Instrument Co., Ltd
Home>Products>DWS-803A Precision Sodium Meter
DWS-803A Precision Sodium Meter
DWS-803A is specifically designed for laboratory ppb level sodium ion measurement electrodes and reference electrodes. Its static measurement performa
Product details

TECHNICAL INDEX

1、 Display mode: Dual row digital display simultaneously

2、 Measurement range: 0.2ug/L to 23000mg/L, 0.00 Pa to 8.00 Pa- 1999~+1999 mV, 0~99.9℃

3、 Resolution: 0.1ug/L; 0.1 mV,0.1℃

4、 Repeatability: ± 2.5% of the reading

5、 Accuracy: ± 2.5% of reading, ± 0.2 mV, ± 0.3 ℃

6、 Continuous stability time:>100 seconds

7、 Detection limit: 0.2ug/L (8pNa)

8、 Input impedance:> ten12Ω

9、 Temperature compensation range: automatic/manual 0-99.9 ℃

10、 Response speed: 32 seconds @ 25 ℃, 50 seconds @ 15 ℃

11、 Communication interface: RS232 bidirectional communication

12、 Power supply: AC220V ± 22V, 50Hz ± 1Hz, external power supply

13、 Working conditions: ambient temperature 0-60 ℃, relative humidity ≤ 90 ℅

14、 Power off storage:>5 years

15、 Clock accuracy: ± 1 minute/month

16、 Calibration methods: one point calibration, two-point calibration, manual input of E0 S

17、 Size and weight: 202x202x68mm, 1.0Kg

Standard configuration

One DWS-51 secondary meter, one HT5 power supply, one 2802 reference and one 2801 measuring electrode, one T22 temperature electrode, and one set of electrode brackets.

main features

Double row digital display simultaneously shows μ g/L/pNa/mV and temperature. Microcomputerization: Using high-performance CPU chips, high-precision AD conversion technology, and SMT surface mount technology, it completes multi parameter measurement, temperature compensation, high accuracy, and good repeatability. High reliability: Single board structure, touch sensitive buttons, no switch knob. Input impedance>1012Ω. High sensitivity: 0.2ppb (8pNa) can be detected in static water samples. Long stable time with a very stable reading. Waterproof and anti-corrosion design. Network function: RS232 bidirectional communication, compatible with printer or TLab1.0 communication and data processing software. Industrial control watchdog: ensures that the instrument does not crash.

1.jpg

TLab1.0 software displays instrument parameters, measured values, and curves